FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4791020 · Received May 22, 2015

Report

Report Number
3004209178-2015-58892
Event Type
Malfunction
Date Received
May 22, 2015
Date of Event
April 15, 2014
Report Date
May 4, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION, PRIME/COMPROMISED FORCE SENSOR SYSTEM, AND DISPLACEMENT TESTS. PUMP WAS RECEIVED STUCK IN A MOTOR ERROR ALARM LOOP DURING BOLUS DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING. PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, A MINOR SCRATCHED LCD WINDOW, AND A SCRATCHED RESERVOIR TUBE WINDOW. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD A MOTOR ERROR ALARM ON THE INSULIN PUMP. CUSTOMER STATED THAT THE ALARM OCCURRED 3 TIMES. INSULIN PUMP WILL BE REPAIRED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335705 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWH

Patients

Seq Age Sex Outcome Treatment
1