FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4791001 · Received May 22, 2015

Report

Report Number
3004209178-2015-58771
Event Type
Malfunction
Date Received
May 22, 2015
Date of Event
May 3, 2015
Report Date
May 3, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

ALL BUTTONS FUNCTIONED PROPERLY DURING TESTING. HOWEVER, THE INSULIN FOUND MOISTURE DAMAGE ON THE KEYPAD TRACES DURING VISUAL INSPECTION. THE INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON LCD WINDOW, AND SCRATCHED RESERVOIR TUBE WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL TO HAVE EXPERIENCED KEYPAD ANOMALY. CUSTOMER REPORTED THAT BUTTONS WERE NOT WORKING CORRECTLY WHEN ATTEMPTS WERE MADE TO CLEAR AN ALARM. CUSTOMER'S BLOOD GLUCOSE WAS 85 MG/DL. CUSTOMER DOES NOT KNOW WHAT CAUSED ANOMALY. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
336190 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-551NAS

Patients

Seq Age Sex Outcome Treatment
1 45 YR