PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2015-58764
- Event Type
- Malfunction
- Date Received
- May 22, 2015
- Date of Event
- May 3, 2015
- Report Date
- May 3, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- PATIENT
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED VIA PHONE, THE INSULIN PUMP ALARMED MOTOR ERROR DURING PRIME. CUSTOMER'S BLOOD GLUCOSE WAS 148 MG/DL. TROUBLESHOOTING WAS PERFORMED. THE DEVICE WAS NOT EXPOSED TO AN MRI AND CUSTOMER DOESN'T USE THE SENSOR FEATURE. CUSTOMER WAS UNABLE TO COMPLETE THE REWIND SEQUENCE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE, REVERT TO BACK UP PLAN, AND THE DEVICE NEED TO BE REPLACED. CUSTOMER AGREED TO RETURN THE DEVICE FOR FURTHER ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 334158 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722LNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |