FLEX-R 'S' #20 25MM
Report
- Report Number
- 2523190-2015-00025
- Event Type
- Injury
- Date Received
- May 20, 2015
- Date of Event
- April 28, 2015
- Report Date
- April 28, 2015
- Manufacturer
- ITEGRA YORK, PA INC.
- Product Code
- EKJ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
TO DATE, THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.
ON 06/15/2015 INTEGRA COMPLETED INVESTIGATION. RESULTS- FAILURE ANALYSIS: DENTAL FILES RETURNED IN USED CONDITION, NOT SHOWING ANY UNUSUAL MARKINGS. THE FILES WERE SHOWING WEAR UPON VISUALLY INSPECTING THE FILES, IT IS NOTICED THAT SOME OF THE FILES ARE BENT AND ONE WITH A BROKEN TIP. QUALITY INSPECTOR TESTED THE BLADES FOR BEND AND TORQUE, AND THE FILES TESTED WITHIN DEFINED SPECIFICATIONS. WITHOUT KNOWING HOW MUCH PRESSURE WAS USED DURING THE PROCEDURE, THE CAUSE OF THE COMPLAINT IS UNDETERMINED. DEVICE HISTORY EVALUATION- NONCONFORMING PRODUCT REPORT/NONCONFORMING MATERIAL REPORT HISTORY: NONE, VARIANCE AUTHORIZATION/DEVIATION HISTORY NONE, ENGINEERING CHANGE ORDER/MANUFACTURING CHANGE ORDER HISTORY: THERE IS NO APPLICABLE ENGINEERING CHANGE ORDER/MANUFACTURING CHANGE ORDER HISTORY, CORRECTIVE ACTION PREVENTIVE ACTION HISTORY: NONE, HEALTH HAZARD EVALUATION HISTORY: NONE. CONCLUSION: THE COMPLAINT REPORT CANNOT BE CONFIRMED, DUE TO THE INSTRUMENT TESTING WITHIN DEFINED SPECIFICATION.
CUSTOMER INITIALLY REPORTS THAT 2 FILES BROKE OFF IN PATIENTS MOUTHS. PIECE COULD NOT BE REMOVED FROM ROOT CANAL, NO ILL EFFECTS SEEN. NUMBER 2 OF 2 RELATED COMPLAINTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 328814 | FLEX-R 'S' #20 25MM | M50-ENDODONTICS | EKJ | ITEGRA YORK, PA INC. | 19052 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Other |