FDA Adverse Event Injury Summary report: N

FLEX-R 'S' #20 25MM

MDR report key: 4790480 · Received May 20, 2015

Report

Report Number
2523190-2015-00025
Event Type
Injury
Date Received
May 20, 2015
Date of Event
April 28, 2015
Report Date
April 28, 2015
Manufacturer
ITEGRA YORK, PA INC.
Product Code
EKJ
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TO DATE, THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.

Additional Manufacturer Narrative · 1

ON 06/15/2015 INTEGRA COMPLETED INVESTIGATION. RESULTS- FAILURE ANALYSIS: DENTAL FILES RETURNED IN USED CONDITION, NOT SHOWING ANY UNUSUAL MARKINGS. THE FILES WERE SHOWING WEAR UPON VISUALLY INSPECTING THE FILES, IT IS NOTICED THAT SOME OF THE FILES ARE BENT AND ONE WITH A BROKEN TIP. QUALITY INSPECTOR TESTED THE BLADES FOR BEND AND TORQUE, AND THE FILES TESTED WITHIN DEFINED SPECIFICATIONS. WITHOUT KNOWING HOW MUCH PRESSURE WAS USED DURING THE PROCEDURE, THE CAUSE OF THE COMPLAINT IS UNDETERMINED. DEVICE HISTORY EVALUATION- NONCONFORMING PRODUCT REPORT/NONCONFORMING MATERIAL REPORT HISTORY: NONE, VARIANCE AUTHORIZATION/DEVIATION HISTORY NONE, ENGINEERING CHANGE ORDER/MANUFACTURING CHANGE ORDER HISTORY: THERE IS NO APPLICABLE ENGINEERING CHANGE ORDER/MANUFACTURING CHANGE ORDER HISTORY, CORRECTIVE ACTION PREVENTIVE ACTION HISTORY: NONE, HEALTH HAZARD EVALUATION HISTORY: NONE. CONCLUSION: THE COMPLAINT REPORT CANNOT BE CONFIRMED, DUE TO THE INSTRUMENT TESTING WITHIN DEFINED SPECIFICATION.

Description of Event or Problem · 1

CUSTOMER INITIALLY REPORTS THAT 2 FILES BROKE OFF IN PATIENTS MOUTHS. PIECE COULD NOT BE REMOVED FROM ROOT CANAL, NO ILL EFFECTS SEEN. NUMBER 2 OF 2 RELATED COMPLAINTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
328814 FLEX-R 'S' #20 25MM M50-ENDODONTICS EKJ ITEGRA YORK, PA INC. 19052

Patients

Seq Age Sex Outcome Treatment
1 51 YR Other