FDA Adverse Event Malfunction Summary report: N

PERFECTO2 V WITH SENSOR 9153650799

MDR report key: 4790255 · Received May 22, 2015

Report

Report Number
1031452-2015-13891
Event Type
Malfunction
Date Received
May 22, 2015
Report Date
April 30, 2015
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

NO END USER INFORMATION PROVIDED. THE PRODUCT WAS EVALUATED AND REPAIRED BY AN INDEPENDENT REPAIR CENTER. A FOLLOW-UP WILL BE SENT IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

PER THE INDEPENDENT REPAIR CENTER, THE CUSTOMER ALLEGED PROBLEM IS ALARMING/RED LIGHT/ALARM WILL NOT FUNCTION. THE KEY FAILURE IS THE MANIFOLD VALVE IS LEAKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
334311 PERFECTO2 V WITH SENSOR 9153650799 GENERATOR, OXYGEN, PORTABLE CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other