FDA Adverse Event Malfunction Summary report: N

PERFECTO2 OXYGEN CONCENTRATOR WITH SENSO2 9153646963

MDR report key: 4785320 · Received May 20, 2015

Report

Report Number
1031452-2015-13832
Event Type
Malfunction
Date Received
May 20, 2015
Report Date
April 30, 2015
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

DEVICE RETURNED FOR EVALUATION. FAILURE MODES: 4-WAY VALVE 1101141 STUCK, MANIFOLD (B)(4).

Additional Manufacturer Narrative · 1

NO END USER INFORMATION PROVIDED. A FOLLOW-UP WILL BE SENT IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DEALER ALLEGES UNIT HAS A LOCKED COMPRESSOR. RED LIGHT AND ALARM KICKS THE RESET BUTTON. INTERMITTENT START UP ALARM. NO FURTHER INFORMATION PROVIDED BY DEALER.

Description of Event or Problem · 1

DEALER ALLEGES UNIT HAS A LOCKED COMPRESSOR. RED LIGHT AND ALARM KICKS THE RESET BUTTON. INTERMITTENT START UP ALARM. NO FURTHER INFORMATION PROVIDED BY DEALER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
329220 PERFECTO2 OXYGEN CONCENTRATOR WITH SENSO2 9153646963 GENERATOR, OXYGEN, PORTABLE CAW INVACARE FLORIDA OPERATIONS IRC5PO2