HALLU FIX
Report
- Report Number
- 478433
- Event Type
- Injury
- Date Received
- August 14, 2003
- Date of Event
- August 6, 2003
- Report Date
- August 6, 2003
- Manufacturer
- WRIGHT MEDICAL CO
- Product Code
- HWC
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MN, US
- Reporter Occupation
- NURSE
Narratives
ADD'L INFO REC'D FROM MFR 10/1/03: WRIGHT MEDICAL TECHNOLOGY, INC IS NOT THE MFR OF THIS PRODUCT BUT THE DISTRIBUTOR. THE PRODUCT IS MFG BY NEW DEAL, INC. THUS WRIGHT IS FORWARDING A COPY OF LETTER TO THE NEW DEAL OFFICE LOCATED AT: NEW DEAL, INC., 2317 BRASSINGTON LANE, PLANO, TX 75075.
SURGEON WAS INSERTING SCREW WHILE PERFORMING OSTEOTOMY OF 5TH METATARSAL. SCREW WAS A HALLUFIX SNAP OFF SCREW THAT THE TOP PORTION FAILED TO SNAP OFF. IN THE PROCESS OF SNAPPING THE SCREW TOP OFF, THE BONE CRACKED. THE SURGICAL PROCEDURE REQUIRED DIFFERENT REPAIR THAN WAS PLANNED.
ADD'L INFO REC'D FROM MFR 10/03/03: MFR HAS BEEN INFORMED THAT WRIGHT MEDICAL TECHNOLOGIES RECEIVED A COPY OF USER FACILITY REPORT UNDER THE MEDICAL DEVICE REPORTING REGULATION (21 CFR 803). THIS MDR IS DUE TO A NEWDEAL'S PRODUCT, BRAND: HALLUFIX, GENERIC: HALLUFIX SNAP-OFF SCREW, PRODUCT CODE: SCREW, FIXATION, BONE. CATALOG NUMBER: 117 012 AND LOT NUMBER B206. MFR HAS NOT PREVIOUSLY SUBMITTED A MEDWATCH REPORT FOR THIS EVENT. MFR WOULD LIKE TO ADD MORE INFO ABOUT THE PRODUCT AND THE SURGICAL TECHNIQUE. FOR THE 5TH METATARSAL OSTEOSYNTHESIS MFR THINKS THAT THE DIAMETER, THE LENGTH AND THE TORQUE OF THIS PRODUCT ARE NOT ADAPTED WITH THE SURGICAL PROCEDURE. IN THIS CASE OTHER PRODUCTS WITH THINNER DIAMETER ARE MORE ADAPTED WITH THE BONE SIZE. IN CASE OF THE SURGEON WOULD LIKE TO USE THIS SURGICAL PROCEDURE, SURGEON SHOULD TAKE CARE OF THE SNAP-OFF SCREW AND ADAPT THE SURGICAL GESTURE BY CUTTING THE SNAP-OFF COMPONENT WITH A FORCEPS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HALLU FIX | HALLU FIX SNAP OFF SCREW | HWC | WRIGHT MEDICAL CO | B206 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Required Intervention |