FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4781850 · Received May 19, 2015

Report

Report Number
2531779-2015-16556
Event Type
Injury
Date Received
May 19, 2015
Report Date
May 11, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION (B)(6) 2015 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/16/2015 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX (BB) DATA INDICATES A TEMP BASAL WAS ACTIVE ON (B)(6) 2015. THE TEMP BASAL WAS CANCELLED WHEN A PRIME WAS PERFORMED AT 12:54. THE BB DOES NOT SHOW A TEMP BASAL PROGRAM RUNNING @ TIME OF THE COMPLAINT. A TEMP BASAL WAS PROGRAMMED; PUMP LOCK FEATURE WAS ENABLED AND DISABLED WITH NO CHANGE OBSERVED TO THE PROGRAMMED TEMP BASAL. THERE WERE NO EAW'S IN THE ALARM HISTORY/BB ASSOCIATED WITH A TEMP BASAL CANCELLATION. INVESTIGATORS WERE UNABLE TO CONFIRM/DUPLICATE COMPLAINT OF TEMP BASAL CANCELLED WHEN PUMP IS LOCKED/UNLOCKED. THERE WAS NO DEFECT FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(6) 2015 ALLEGING A OTHER (UNUSUAL) ISSUE. THE REPORTER STATED THAT THE PATIENT HAD A BLOOD GLUCOSE (BG) OF OVER 27.7MMOL/L WITHOUT SYMPTOMS. THE PATIENT DID NOT RECEIVE ANY TREATMENT ABOVE AND BEYOND THE USUAL ROUTINE CARE OF DIABETES MANAGEMENT. CUSTOMER SUPPORT (CS) COMPLETED TROUBLESHOOTING AND THE TEMP BASAL WAS CANCELLED AFTER PUMP LOCKING. THE PATIENT WAS INSISTENCE AND LOST CONFIDENCE. THIS REPORT IS BEING MADE DUE TO THE BG EXCURSION THE PATIENT EXPERIENCED DUE TO UNUSUAL ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
325707 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 11 YR Life Threatening