MONOPTY DISPOSABLE CORE BIOPSY INSTRUMENT
Report
- Report Number
- 2020394-2015-00488
- Event Type
- Malfunction
- Date Received
- April 16, 2015
- Report Date
- March 26, 2015
- Manufacturer
- BARD PERIPHERAL VASCULAR, INC.
- Product Code
- KNW
- PMA / PMN Number
- K922939
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE LOT NUMBER HAS BEEN PROVIDED AND THE LOT DEVICE HISTORY RECORDS HAVE BEEN REVIEWED THE LOT MET ALL RELEASE CRITERIA. THERE WAS NOTHING FOUND TO INDICATE THERE WAS A MFG RELATED CAUSE FOR THIS EVENT. THE DEVICE WAS RETURNED HALF PRIMED, WITH THE CANNULA RETRACTED, EXPOSING THE SAMPLE NOTCH. LOOSE PARTICLES COULD BE HEARD WITHIN THE DEVICE. THERE WERE NO VISUAL ANOMALIES NOTED ON THE SAMPLE. THE SAMPLE WAS TESTED BY ATTEMPTING TO TWIST THE ROTATIONAL KNOB ONCE HOWEVER THE DEVICE WOULD NOT PRIME. FURTHER FUNCTIONAL TESTING COULD NOT BE PERFORMED AS THE DEVICE COULD NOT BE FULLY PRIMED. THE SAMPLE WAS DISASSEMBLES AND THE INTERNAL COMPONENTS WERE EXAMINED. THE CANNULA HUB WAS FOUND TO BE BROKEN. THE INVESTIGATION IS CONFIRMED FOR A BREAK, AS THE CANNULA HUB WAS FOUND TO BE BROKEN. THE INVESTIGATION IS INCONCLUSIVE FOR FAILURE TO FIRE, AS THE DEVICE COULD NOT BE FUNCTIONALLY TESTED DUE TO THE BROKEN COMPONENTS. THE ROOT CAUSE FOR THIS EVENT WAS DETERMINED TO BE SUPPLIER RELATED DUE TO OVER-PROCESSED RESIN.
IT WAS REPORTED THAT THE PROBE FAILED TO FIRE DURING THE US GUIDED BREAST BIOPSY AFTER BEING FULLY PRIMED. A COAXIAL WAS USED DURING THE PROCEDURE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256272 | MONOPTY DISPOSABLE CORE BIOPSY INSTRUMENT | KNW | BARD PERIPHERAL VASCULAR, INC. | REYI2147 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |