FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 4781488 · Received May 19, 2015

Report

Report Number
1416980-2015-21978
Event Type
Malfunction
Date Received
May 19, 2015
Date of Event
December 30, 2014
Report Date
April 24, 2015
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS MANUFACTURED DURING JANUARY 2014. THE DEVICE WAS RECEIVED AND THE EVALUATION IS COMPLETE. DURING THE EVENT HISTORY LOG REVIEW, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THE HOMECHOICE DEVICE RECEIVED A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDED FUNCTIONAL AND ELECTRICAL TESTING OF THE DEVICE. A DEVICE HISTORY RECORD REVIEW REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE ISSUE. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THIS ISSUE WAS UNDETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN 105 (NIGHT DRAIN #5) ALARM WAS IDENTIFIED IN THE LOG. THIS ALARM INDICATES THAT THE PATIENT DRAINED GREATER THAN OR EQUAL TO 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME. THE ALARM OCCURRED ON (B)(6) 2014 AT 02:31:09. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
324775 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 43 YR