FDA Adverse Event
Malfunction
Summary report: N
ANIMAS VIBE
MDR report key: 4781181
·
Received May 19, 2015
Report
- Report Number
- 2531779-2015-16523
- Event Type
- Malfunction
- Date Received
- May 19, 2015
- Report Date
- May 5, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6)2015 WITH THE FOLLOWING FINDINGS: INVESTIGATION REVEALED A CRACK IN THE BATTERY COMPARTMENT. UNRELATED TO THIS ISSUE, INVESTIGATION REVEALED THAT THE AUDIO BOLUS BUTTON WAS MISSING FROM THE PUMP. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A CRACK IN THE BATTERY COMPARTMENT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2015.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 325959 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|