SPECTRUM INFUSION PUMP
MDR report key: 4781082
·
Received May 15, 2015
Report
- Report Number
- 1314492-2015-06003
- Event Type
- Malfunction
- Date Received
- May 15, 2015
- Date of Event
- April 1, 2015
- Report Date
- April 22, 2015
- Manufacturer
- BAXTER HEALTHCARE CORP.
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE SYSTEM ERROR 105, WHICH WAS OBSERVED DURING POWER UP. SYSTEM ERROR 105 WAS CONFIRMED AND CAUSED BY A FAILED MOTOR FLEX. THE FAILED MOTOR ASSEMBLY WAS REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 105. IT WAS ALSO REPORTED THERE WAS NO PT INVOLVEMENT. THE EVENT OCCURRED IN THE ONCOLOGY CARE AREA.