FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4780883 · Received May 19, 2015

Report

Report Number
3004209178-2015-57556
Event Type
Malfunction
Date Received
May 19, 2015
Date of Event
April 28, 2015
Report Date
April 28, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, AND EXCESSIVE NO DELIVERY ALARM TESTS. NO MOTOR ERROR ALARM WAS NOTED AND THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP WAS RECEIVED WITH CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP, CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS AND A CRACKED DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE CALLER DID NOT KNOW THE BLOOD GLUCOSE READING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
326292 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554WWL

Patients

Seq Age Sex Outcome Treatment
1