FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4780645
·
Received May 19, 2015
Report
- Report Number
- 3004209178-2015-57352
- Event Type
- Malfunction
- Date Received
- May 19, 2015
- Date of Event
- December 18, 2014
- Report Date
- April 28, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FINDINGS: CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND BROKEN OFFEND CAP NOTED DURING VISUAL INSPECTION. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
Description of Event or Problem · 1
A CUSTOMER REPORTED VIA PHONE CALL INDICATING PHYSICAL DAMAGE IN THE FORM OF CRACKS ON THEIR INSULIN PUMP. THE CUSTOMER SENT THE PRODUCT BACK FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 325117 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-554WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |