FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4780645 · Received May 19, 2015

Report

Report Number
3004209178-2015-57352
Event Type
Malfunction
Date Received
May 19, 2015
Date of Event
December 18, 2014
Report Date
April 28, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FINDINGS: CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND BROKEN OFFEND CAP NOTED DURING VISUAL INSPECTION. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

A CUSTOMER REPORTED VIA PHONE CALL INDICATING PHYSICAL DAMAGE IN THE FORM OF CRACKS ON THEIR INSULIN PUMP. THE CUSTOMER SENT THE PRODUCT BACK FOR ANALYSIS.