FDA Adverse Event Malfunction Summary report: N

ELIX

MDR report key: 4780577 · Received May 12, 2015

Report

Report Number
4780577
Event Type
Malfunction
Date Received
May 12, 2015
Date of Event
May 6, 2015
Report Date
May 11, 2015
Manufacturer
EMD MILLIPORE CORPORATION
Product Code
FIP
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MO, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

BEGINNING WITH THE START UP IN THE AM FOR MIDNIGHT SHIFT, WE STARTED HAVING QC ISSUES WITH CALCIUMS ON THE CE TO THE POINT WE COULD NOT GET THE QC TO WORK AT ALL. ROCHE HOTLINE WAS CONTACTED AND MULTIPLE PROCEDURES (CLEANS, FLUSHES, RECAL'S, REAGENT DUMPS, ETC.) WERE TRIED. NOTHING FIXED IT. WHEN DAY SHIFT ATTEMPTED THEIR START UP ON THE OTHER ANALYZER (CEE), WE EXPERIENCED THE SAME ISSUE. WHEN WE CALLED SERVICE, WE WERE TOLD THAT THE WATER COMING INTO THE ANALYZERS COULD BE A POTENTIAL CAUSE FOR THE PROBLEM. SERVICE DID COME IN AND DUMPED AND CLEANED THE WATER BATH ON THE CEE, AND CHECKED FOR ANYTHING ON THE INSTRUMENT THAT COULD HAVE CAUSED THE ISSUE. BUT WITH THE PROBLEM OCCURRING ON BOTH ANALYZERS, WE FELT STRONGLY IT WAS THE WATER THAT HAD CAUSED THE PROBLEMS. WE HAD TO DUMP ALL CALCIUM PACKS THAT HAD BEEN LOADED DURING THAT TIME, AS THEY ALL HAD USED WATER FOR DILUTING REAGENTS. BY THE END OF DAY, WE WERE FINALLY ABLE TO GET QC TO WORK AND START REPORTING PTS AGAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310806 ELIX PURIFIER, WATER (ABSORPTION, DEIONIZATION, MEMBRANE FILTER, FIP EMD MILLIPORE CORPORATION ELIX 100 (CUSTOM) F3MA66536
310807 ELIX PURIFIER, WATER (ABSORPTION, DEIONIZATION, MEMBRANE FILTER, FIR EMD MILLIPORE CORPORATION ELIX 100 (CUSTOM) F3MA66536
310808 COBAS 501E TITRIMETRIC WITH EDTA AND INDICATOR, CALCIUM JFP ROCHE DIAGNOSTICS GMBH CE (C501) 609429
310822 COBAS 501E TITRIMETRIC WITH EDTA AND INDICATOR, CALCIUM JFP ROCHE DIAGNOSTICS GMBH CEE (C501) 609429

Patients

Seq Age Sex Outcome Treatment
1 * WE HAVE A ROCHE CEE AND CE, AND EVERYTHING WAS IN| LINE AND BOTH WERE IN USE.| USE. THE EVOQUA HAS 2 SYSTEMS SUPPLYING THE ROCHE