SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2015-21917
- Event Type
- Malfunction
- Date Received
- May 19, 2015
- Date of Event
- May 3, 2015
- Report Date
- May 3, 2015
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE PATIENT LINE EXTENSION CLAMP BEING CLOSED DURING PRIME IS A KNOWN CAUSE OF THIS ALARM. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE INSTRUCTS THE USER TO OPEN THE CLAMP ON THE PATIENT LINE TO ENSURE PROPER PRIMING. IT PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET) ALARM. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING THE INITIAL DRAIN OF PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING, IT WAS DISCOVERED THAT THE CLAMP ON THE PATIENT LINE EXTENSION WAS CLOSED DURING PRIME. THE PATIENT STATED THEY WOULD START OVER WITH NEW SUPPLIES AND THE TECHNICAL SERVICE REPRESENTATIVE EXPLAINED THE PROPER PRIMING PROCEDURE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 326171 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | HOMECHOICE |