FDA Adverse Event Injury Summary report: N

BONE PIN- 3.2 X 140MM- STERILE (2 PACK)

MDR report key: 4778568 · Received May 18, 2015

Report

Report Number
3005985723-2015-00077
Event Type
Injury
Date Received
May 18, 2015
Date of Event
May 7, 2015
Report Date
May 7, 2015
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K112507
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

REPORTED EVENT: DURING PARTIAL KNEE REPLACEMENT SURGERY, THE TIP OF A 3.2X140MM BONE PIN BROKE OFF AND EMBEDDED IN PATIENT AS SURGEON ATTEMPTED DRILL THE PIN INTO THE PATIENT. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION WAS NOT PERFORMED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. DIMENSION INSPECTION COMPLETED AT THE TIME OF MANUFACTURING SHOWS THE LOT WAS WITHIN SPECIFICATION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS FOR THE ASSOCIATED LOT INDICATED (B)(4) DEVICES (111621 NON-STERILE BONE PIN, SINGLE) WERE MANUFACTURED AND SHIPPED TO (B)(6) ON 03/12/2015 AND ACCEPTED INTO FINAL STOCK ON 03/12/2015 PER ERP. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN TRACKWISE AND CATSWEB RELATED TO P/N 111620, LOT NUMBER W39911-2 SHOWS NO COMPLAINTS RELATED TO THE FAILURE. (B)(4). CONCLUSIONS: AS PART OF THE INVESTIGATION INTO THIS COMPLAINT, THE STRYKER REP THAT WAS PRESENT FOR THE CASE WAS CONTACTED. IT WAS DETERMINED THAT THE SURGEON DID NOT USE THE DRILL GUIDE DURING PLACEMENT OF BOTH BONE PINS. DUE TO THE NON-USE OF A DRILL GUIDE AS REQUIRED PER MAKOPLASTY PARTIAL KNEE APPLICATION USER GUIDE, 206388 REVISION 01 FOR PLACEMENT OF BONE PINS, THIS EVENT CONSTITUTES AN OFF-LABEL USE OF THE DEVICE. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD. DEVICE WAS NOT RETURNED FOR EVALUATION.

Description of Event or Problem · 1

THE SURGEON PERFORMED A PARTIAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2015 USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS MULTICOMPARTMENTAL KNEE (MCK) IMPLANTS. DURING THE CASE, THE TIP OF THE BONE PIN BROKE OFF IN THE PATIENT'S TIBIA. THE SURGEON DECIDED TO LEAVE THE TIP OF THE BONE PIN IN THE PATIENT'S TIBIA. THE OUTCOME OF THE CASE WAS SUCCESSFUL.

Description of Event or Problem · 1

THE SURGEON PERFORMED A PARTIAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2015 USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS MULTICOMPARTMENTAL KNEE (MCK) IMPLANTS. DURING THE CASE, THE TIP OF THE BONE PIN BROKE OFF IN THE PATIENT'S TIBIA. THE SURGEON DECIDED TO LEAVE THE TIP OF THE BONE PIN IN THE PATIENT'S TIBIA. THE OUTCOME OF THE CASE WAS SUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321727 BONE PIN- 3.2 X 140MM- STERILE (2 PACK) STEREOTAXIC DEVICE, ACCESSORY OLO MAKO SURGICAL CORP. W39911-2

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention