HEARTWARE VENTRICULAR ASSIST SYSTEM
Report
- Report Number
- 3007042319-2015-00877
- Event Type
- Malfunction
- Date Received
- May 15, 2015
- Date of Event
- April 15, 2015
- Report Date
- April 16, 2015
- Manufacturer
- HEARTWARE, INC
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Removal / Correction Number
- Z-1607-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. THE INSTRUCTIONS FOR USE AND PATIENT MANUAL INCLUDE A REFERENCE GUIDE FOR BOTH VISUAL AND TONE ALARMS AND ASSOCIATED CAUSES AND ACTIONS TO TAKE. ADDITIONALLY THERE IS A WARNING TO KEEP SPARE, FULLY CHARGED BATTERIES AND BACK UP CONTROLLER AVAILABLE AT ALL TIME. THE STEPS FOR EXCHANGE OF DEVICES ARE ALSO OUTLINED. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT WHEN NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. THE COMPANY IS SEEKING THIS INFORMATION THROUGH THE EVENT INVESTIGATION. THIS IS ONE OF SIX REPORTS (3007042319-2015-0000876,3007042319-2015-0000877,3007042319-2015-0000878,3007042319-2015-0000879,3007042319-2015-0000880 AND 3007042319-2015-0000881,) SUBMITTED FOR DEVICES RELATED TO THE SAME EVENT. BATTERIES HAVE NOT BEEN RECEIVED.
BASED ON THE RESULTS OF THE PREVIOUS MANUFACTURER'S INTERNAL INVESTIGATION, FIELD ACTIONS AND CHANGES TO THE BATTERY INTERNAL CELL SUPPLIER, NO ADDITIONAL INVESTIGATION OF THIS EVENT IS REQUIRED AT THIS TIME. THE MOST LIKELY CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO FAULTY INTERNAL CELLS WITHIN THE BATTERY PACK. THIS IS ONE OF SIX REPORTS (3007042319-2015-0000876,3007042319-2015-0000877,3007042319-2015-0000878,3007042319-2015-0000879,3007042319-2015-0000880 AND 3007042319-2015-0000881,) SUBMITTED FOR DEVICES RELATED TO THE SAME EVENT.
THE VENTRICULAR ASSIST DEVICE (VAD) COORDINATOR STATES THAT "ALL SIX BATTERIES SWITCH BACK AND FORTH BEFORE DEPLETING CHARGE." THIS IS REPORT 2 OF 6.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 319507 | HEARTWARE VENTRICULAR ASSIST SYSTEM | CIRCULATORY ASSIST SYSTEM | DSQ | HEARTWARE, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |