FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4771439 · Received May 14, 2015

Report

Report Number
2531779-2015-16030
Event Type
Malfunction
Date Received
May 14, 2015
Report Date
April 30, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 07/25/2016 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/07/2016 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE KEYPAD WAS FOUND TO BE LIFTED. DURING TESTING, ALL OF THE KEYPAD BUTTONS RESPONDED APPROPRIATELY. THE KEYPAD COVER WAS REMOVED AND NO DAMAGE TO THE BUTTON CONTACTS OR KEYPAD FLEX CABLE WAS OBSERVED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(4).

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2015 THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD ISSUE. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
316058 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 12 YR