ANIMAS VIBE
Report
- Report Number
- 2531779-2015-16027
- Event Type
- Malfunction
- Date Received
- May 14, 2015
- Report Date
- April 30, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP #1: DATE OF SUBMISSION 06/30/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/16/2015 WITH THE FOLLOWING FINDINGS: REVIEW OF THE PUMP ALARM HISTORY REVEALED SEVERAL RECORDS OF CALL SERVICE ALARM (012-0195). ON INVESTIGATION, EZ-PRIME WAS PERFORMED SUCCESSFULLY WITHOUT CALL SERVICE ALARM, WARNING, ERROR OR ALARM. THE PUMP WAS EXERCISED FOR 24 HOURS AND FOUND TO BE OPERATING WITHIN THE REQUIRED SPECIFICATION WITHOUT MALFUNCTION. THE PUMP WAS OPENED FOR FURTHER INVESTIGATION AND DID NOT REVEAL ANY DAMAGE, DEFECT OR CONTAMINATION OF THE PUMP¿S INTERIOR COMPONENTS. INVESTIGATION DID NOT DUPLICATE THE COMPLAINT. UNRELATED TO THE ORIGINAL COMPLAINT, THE DISPLAY WAS NOTED TO BE DIM/FADED/DISCOLORED AND THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED BELOW THE BUMPER PAD.
ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 012) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 316007 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |