FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4771357 · Received May 14, 2015

Report

Report Number
3004753838-2015-03810
Event Type
Malfunction
Date Received
May 14, 2015
Date of Event
April 16, 2015
Report Date
April 16, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT TRANSMITTER DEVICE WAS NOT RETURNED TO DEXCOM. THE RECEIVER DEVICE ((B)(4)/LOT NUMBER 5191732), BEING USED WITH THE COMPLAINT TRANSMITTER DEVICE, WAS RETURNED ON 05/07/2015. THE RETURNED COMPLAINT RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. A REVIEW OF THE DOWNLOADED RECEIVER LOG CONFIRMED THE REPORTED EVENT OF A PERMANENT OUT OF RANGE SIGNAL. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT A PERMANENT OUT OF RANGE SIGNAL ON (B)(6) 2015. PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
315948 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. 9438-05 5191459

Patients

Seq Age Sex Outcome Treatment
1 66 YR