G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-03810
- Event Type
- Malfunction
- Date Received
- May 14, 2015
- Date of Event
- April 16, 2015
- Report Date
- April 16, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4). SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT TRANSMITTER DEVICE WAS NOT RETURNED TO DEXCOM. THE RECEIVER DEVICE ((B)(4)/LOT NUMBER 5191732), BEING USED WITH THE COMPLAINT TRANSMITTER DEVICE, WAS RETURNED ON 05/07/2015. THE RETURNED COMPLAINT RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. A REVIEW OF THE DOWNLOADED RECEIVER LOG CONFIRMED THE REPORTED EVENT OF A PERMANENT OUT OF RANGE SIGNAL. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED.
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT A PERMANENT OUT OF RANGE SIGNAL ON (B)(6) 2015. PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 315948 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9438-05 | 5191459 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |