FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4771246
·
Received May 14, 2015
Report
- Report Number
- 3004209178-2015-56184
- Event Type
- Malfunction
- Date Received
- May 14, 2015
- Date of Event
- January 23, 2015
- Report Date
- April 23, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED VIA PHONE CALL THAT HE IS HAVING AN ISSUE WITH HIS PUMP TURNING OFF AND REBOOTING ITSELF. CUSTOMER STATED THAT THE PUMP THEN STARTS COUNTING UP. CUSTOMER STATED THAT THE PUMP WILL GO BLANK. CUSTOMER REPORTED THAT THE BATTERY COMPARTMENT WAS NOT DAMAGED OR CORRODED. TROUBLESHOOTING WAS PERFORMED AND THE CUSTOMER WILL BE SENT A REPLACEMENT BATTERY CAP FOR THE BLANK DISPLAY ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 315312 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |