PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2015-55985
- Event Type
- Malfunction
- Date Received
- May 14, 2015
- Date of Event
- April 21, 2015
- Report Date
- April 22, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
NO BUTTON ERROR ALARM WAS NOTED. HOWEVER, INTERMITTENT BUTTON RESPONSE WAS NOTED DUE TO CORRODED KEYPAD TRACES. NO DISPLAY ANOMALY WAS NOTED. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT A DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP AND A STAINED END CAP STICKER.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED FOR A BUTTON ERROR AND THE KEYPAD BECAME UNRESPONSIVE. IT WAS ALSO REPORTED THAT THE SCREEN FLICKERED AND CHANGED WITHOUT INPUT. THE BLOOD GLUCOSE READING WAS 8.7 MMOL/L. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.