FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4771026 · Received May 14, 2015

Report

Report Number
2032227-2015-15657
Event Type
Malfunction
Date Received
May 14, 2015
Date of Event
April 16, 2015
Report Date
April 23, 2015
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NM
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED ALL FUNCTIONAL TESTS. THE INSULIN PUMP FUNCTIONED PROPERLY. HOWEVER, THE INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE NOTED DUE TO FLATTENED DOME SWITCH ON THE ESC BUTTON. THE INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE NOTED DUE TO FLATTENED DOME SWITCH ON THE ESC BUTTON. THE INSULIN PUMP ALSO HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED RESERVOIR TUBE LIP AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

THE CUSTOMER CALLED AND REPORTED HE WAS HOSPITALIZED WITH BLOOD GLUCOSE OF 29 MG/DL, AFTER THE CUSTOMER PASSED OUT AND HAD A LOW BLOOD GLUCOSE OF 25 MG/DL. THE CUSTOMER REPORTEDLY TREATED WITH FOOD. CUSTOMER STATED THE REASON FOR THE LOW BLOOD GLUCOSE WAS THAT THE INSULIN PUMP BUTTONS MALFUNCTIONED AND INSULIN WAS ACCIDENTALLY PUMPED. CUSTOMER FURTHER STATED HE HAD TO SMACK THE INSULIN PUMP ON THE TABLE TO GET IT TO WORK. NO SIGNIFICANT EVENTS WERE RECALLED LEADING UP TO THE KEYPAD ISSUES. IT WAS ADVISED THAT THE CUSTOMER DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACK-UP PLAN. CUSTOMER WAS FURTHER ADVISED THE INSULIN PUMP WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
315246 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAB

Patients

Seq Age Sex Outcome Treatment
1 51 YR Hospitalization