FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4771022 · Received May 14, 2015

Report

Report Number
3004209178-2015-56111
Event Type
Malfunction
Date Received
May 14, 2015
Date of Event
December 9, 2014
Report Date
April 23, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNIT PASSED FUNCTIONAL TEST INCLUDING DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, AND EXCESSIVE NO DELIVERY ALARM TEST. UNIT FUNCTIONED PROPERLY. HOWEVER NOISY MOTOR NOTED DURING TESTING DUE TO FAULTY MOTOR. UNIT RECEIVED WITH MINOR SCRATCHED LCD WINDOW, CRACKED RESERVOIR TUBE LIP, AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

IT WAS REPORTED VIA PHONE CALL, THAT THE CUSTOMER'S INSULIN PUMP MAKES A GRINDING NOISE. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME WAS UNKNOWN. TROUBLESHOOTING WAS DECLINED. NO SIGNIFICANT EVENT LEADING UP TO THE INCIDENT WAS MENTIONED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
316674 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 62 YR