FDA Adverse Event Malfunction Summary report: N

CAPTURE-R READY INDICATOR RED CELLS

MDR report key: 4769834 · Received May 13, 2015

Report

Report Number
1034569-2015-00078
Event Type
Malfunction
Date Received
May 13, 2015
Date of Event
April 16, 2015
Report Date
May 13, 2015
Manufacturer
IMMUCOR, INC.
Product Code
KSZ
PMA / PMN Number
BK020053
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

IMPLICATED PRODUCT CAPTURE-R READY INDICATOR RED CELL, LOT 221353, EXPIRED PRIOR TO BEING REFERRED TO PI. PRODUCT WAS NOT USED TO CONFIRM REACTIVITY OF D NOR A C ANTIGEN WHILE IN DATE. DHR REVIEW PERFORMED FOR CRRIRC LOT 221353 FOR PERFORMANCE ON THE ECHO PRIOR TO RELEASE. QUALITY CONTROL - FINAL RELEASE: ECHO INSTRUMENTATION: PERFORMED ON (B)(6) 2015. SCREEN ASSAY PERFORMED; CONTROLS RESULTED AS EXPECTED. WBCORQC TUBE 2 CONTAINS ANTI D. IT WAS 4+ POSITIVE. FOUR REPLICATES OF CAPTURE-R NEGATIVE CONTROL, LOT 245150, RESULTED NEGATIVE AS EXPECTED. FOUR REPLICATES OF CAPTURE-R POSITIVE CONTROL (WEAK), LOT 244310, RESULTED POSITIVE AS EXPECTED. FOUR REPLICATES OF KNOWN ANTIBODY NEGATIVE (INERT) IN-HOUSE DONORS RESULTED NEGATIVE AS EXPECTED. DOCUMENTATION STATES THAT ALL WELLS WERE VISUALLY VERIFIED. CAPTURE R INDICATOR CELL POTENCY TESTING WAS PERFORMED ON (B)(6) 2015. TITER FOR ANTI-D WAS 1:64. UNABLE TO RULE OUT COMPROMISED VIAL OF CRRIC ((B)(4)). ISSUE WAS RESOLVED WITH NEW VIAL OF CRRIC ((B)(4)) WITH CRRS LOT R584.

Description of Event or Problem · 1

ON (B)(6) 2015, A CUSTOMER REPORTED THAT UNEXPECTED NEGATIVE RESULTS WERE OBTAINED WHEN TESTING A PATIENT SAMPLE WITH CAPTURE-R READY-SCREEN (3) ON THE GALILEO ECHO USING CAPTURE-R READY-INDICATOR RED CELLS (CRRIRC) LOT 221353, WHEN TESTED ON (B)(6) 2015.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
312018 CAPTURE-R READY INDICATOR RED CELLS ANTI-IGG COATED RED BLOOD CELLS KSZ IMMUCOR, INC. 221353

Patients

Seq Age Sex Outcome Treatment
1 81 YR