CAPTURE-R READY INDICATOR RED CELLS
Report
- Report Number
- 1034569-2015-00078
- Event Type
- Malfunction
- Date Received
- May 13, 2015
- Date of Event
- April 16, 2015
- Report Date
- May 13, 2015
- Manufacturer
- IMMUCOR, INC.
- Product Code
- KSZ
- PMA / PMN Number
- BK020053
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
IMPLICATED PRODUCT CAPTURE-R READY INDICATOR RED CELL, LOT 221353, EXPIRED PRIOR TO BEING REFERRED TO PI. PRODUCT WAS NOT USED TO CONFIRM REACTIVITY OF D NOR A C ANTIGEN WHILE IN DATE. DHR REVIEW PERFORMED FOR CRRIRC LOT 221353 FOR PERFORMANCE ON THE ECHO PRIOR TO RELEASE. QUALITY CONTROL - FINAL RELEASE: ECHO INSTRUMENTATION: PERFORMED ON (B)(6) 2015. SCREEN ASSAY PERFORMED; CONTROLS RESULTED AS EXPECTED. WBCORQC TUBE 2 CONTAINS ANTI D. IT WAS 4+ POSITIVE. FOUR REPLICATES OF CAPTURE-R NEGATIVE CONTROL, LOT 245150, RESULTED NEGATIVE AS EXPECTED. FOUR REPLICATES OF CAPTURE-R POSITIVE CONTROL (WEAK), LOT 244310, RESULTED POSITIVE AS EXPECTED. FOUR REPLICATES OF KNOWN ANTIBODY NEGATIVE (INERT) IN-HOUSE DONORS RESULTED NEGATIVE AS EXPECTED. DOCUMENTATION STATES THAT ALL WELLS WERE VISUALLY VERIFIED. CAPTURE R INDICATOR CELL POTENCY TESTING WAS PERFORMED ON (B)(6) 2015. TITER FOR ANTI-D WAS 1:64. UNABLE TO RULE OUT COMPROMISED VIAL OF CRRIC ((B)(4)). ISSUE WAS RESOLVED WITH NEW VIAL OF CRRIC ((B)(4)) WITH CRRS LOT R584.
ON (B)(6) 2015, A CUSTOMER REPORTED THAT UNEXPECTED NEGATIVE RESULTS WERE OBTAINED WHEN TESTING A PATIENT SAMPLE WITH CAPTURE-R READY-SCREEN (3) ON THE GALILEO ECHO USING CAPTURE-R READY-INDICATOR RED CELLS (CRRIRC) LOT 221353, WHEN TESTED ON (B)(6) 2015.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312018 | CAPTURE-R READY INDICATOR RED CELLS | ANTI-IGG COATED RED BLOOD CELLS | KSZ | IMMUCOR, INC. | 221353 |