IPC® HANDPIECE ATTACHMENT - INDIGO¿ DRILL
Report
- Report Number
- 1625507-2015-00020
- Event Type
- Malfunction
- Date Received
- May 11, 2015
- Date of Event
- April 12, 2015
- Report Date
- April 17, 2015
- Manufacturer
- MDT POWERED SURGICAL SOLUTIONS
- Product Code
- HBE
- PMA / PMN Number
- K081475
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES. CONCOMITANT MEDICAL PRODUCTS: 1845020: ATTACHMENT INDIGO OTOL ANGLED, SERIAL # (B)(4), LOT #207262552, MANUFACTURED DATE ¿ AUG/07/2013. 1845020: ATTACHMENT INDIGO OTOL ANGLED, SERIAL #(B)(4), LOT # 207262552, MANUFACTURED DATE ¿AUG/07/2013. 1845000: DRILL INDIGO HIGH SPEED OTOLOGIC, SERIAL # (B)(4). (B)(4). THE THREE ATTACHMENTS WERE RECEIVED AT XOMED FOR PRE-REPAIR EVALUATION. PRE-REPAIR TEMPERATURE TESTING AT ONE MINUTE FOUND THE INDIGO ANGLED ATTACHMENT (PART # 1845020, SN# (B)(4)) MEASURED: NOSE - 75 DEGREES F, MIDDLE - 78 DEGREES F, REAR 81 - DEGREES F. THE COLLET WOULD NOT HOLD THE BUR. OVERHEATING COULD NOT BE DUPLICATED OR CONFIRMED. PRE-REPAIR TEMPERATURE TESTING FOR THE INDIGO ATTACHMENT (PART # 1845010, SN# (B)(4)) COULD NOT BE COMPLETED AS THE ATTACHMENT WOULD NOT SIT RIGHT ON THE DRILL. OVERHEATING COULD NOT BE DUPLICATED OR CONFIRMED. PRE-REPAIR TEMPERATURE TESTING AT ONE MINUTE FOUND THE INDIGO ANGLED ATTACHMENT (PART # 1845020, SN# (B)(4)) MEASURED: NOSE - 74 DEGREES F, MIDDLE - 74 DEGREES F, REAR - 73 DEGREES F. THE COLLET WOULD NOT HOLD THE BUR. THE OVERHEATING COULD NOT BE DUPLICATED OR CONFIRMED. ALL THREE ATTACHMENTS WERE SENT TO FORT WORTH MPSS FOR REPAIR.
FURTHER TESTING FOUND THE FOLLOWING: PRODUCT ID#1845020, SN# (B)(4), LOT# 207262552 ¿ AFTER RUNNING FOR 5 MINUTES THE DEVICE TEMPERATURE MEASURED 136.8 DEGREES F. EVALUATION OF THE DEVICE FOUND THAT THE BEARINGS AND GEARS WERE CORRODED. PRODUCT ID#1845010, SN# (B)(4), LOT# 207255663 ¿ AFTER RUNNING FOR 5 MINUTES, THE DEVICE TEMPERATURE MEASURED 118.6 DEGREES F. EVALUATION OF THE DEVICE FOUND THAT THE BEARINGS WERE WORN. PRODUCT ID#1845020 SN# (B)(4), LOT# 207262552 ¿ AFTER RUNNING FOR 5 MINUTES THE DEVICE TEMPERATURE MEASURED 128.3 DEGREES F. EVALUATION OF THE DEVICE FOUND THAT THE BEARINGS AND GEARS WERE CORRODED.
IT WAS REPORTED THAT THREE ATTACHMENTS WERE OVERHEATING AND CANNOT SECURE THE BUR PROPERLY. LUBRICATION PERFORMED AS INDICATED IN USER'S MANUAL. THERE WAS NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 307173 | IPC® HANDPIECE ATTACHMENT - INDIGO¿ DRILL | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | MDT POWERED SURGICAL SOLUTIONS | 1845010 | 207255663 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |