LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2015-02758
- Event Type
- Injury
- Date Received
- May 6, 2015
- Date of Event
- April 23, 2014
- Report Date
- April 30, 2015
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THERE WAS NO DEATH OR DEVICE MALFUNCTION ASSOCIATED WITH THE INAPPROPRIATE DEFIBRILLATION. DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) AND BELT SN (B)(4) WAS COMPLETED. THE MONITOR AND ELECTRODE BELT WERE FULLY FUNCTIONAL AND ABLE TO DETECT AND TREAT A PATIENT. DEVICE MANUFACTURE DATE: MONITOR SN (B)(4); ELECTRODE BELT SN (B)(4): 01/2013. INAPPROPRIATE DEFIBRILLATIONS ARE AN ANTICIPATED RISK ASSOCIATED WITH THE USE OF THE LIFEVEST. PATIENTS ARE INSTRUCTED THROUGH ALARMS, VOICE MESSAGES, IFU, AND TRAINING TO PRESS THE RESPONSE BUTTONS TO PREVENT AN INAPPROPRIATE DEFIBRILLATION. (B)(4). A SUMMARY OF THE SAFETY AND EFFECTIVENESS DATA (SSED), INCLUDING THE INAPPROPRIATE DEFIBRILLATION SAFETY OBJECTIVE SUPPORTING FDA'S APPROVAL OF THE LIFEVEST, CAN BE FOUND AT HTTP://WWW.ACCESSDATA.FDA.GOV/CDRH_DOCS/PDF/P010030B.PDF.
DURING A RETROACTIVE REVIEW OF PAST TREATMENT EVENTS FOR POTENTIAL ADVERSE EVENTS, CONDUCTED AS A CAPA IN RESPONSE TO A RECENT FDA INSPECTION, A REPORTABLE ADVERSE EVENT WAS DISCOVERED. THE GRAND-DAUGHTER OF A (B)(6) MALE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PATIENT WAS TREATED SEVERAL TIMES. THE PATIENT WAS IN THE HOSPITAL AT THE TIME OF THE TREATMENT. A REVIEW OF THE EVENT INDICATES THAT THE PATIENT EXPERIENCED AN INAPPROPRIATE DEFIBRILLATION EVENT CONSISTING OF 5 INAPPROPRIATE TREATMENTS. RAPID ATRIAL FIBRILLATION AT 152-162 BPM CONTRIBUTED TO THE FALSE DETECTIONS. THE RESPONSE BUTTONS WERE NOT PRESSED DURING THE ENTIRE EVENT. THE PATIENT REMAINED IN THE HOSPITAL AND CONTINUED USE OF THE LIFEVEST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 296757 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Hospitalization| O |