FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 4760296 · Received May 8, 2015

Report

Report Number
1416980-2015-20721
Event Type
Malfunction
Date Received
May 8, 2015
Date of Event
April 1, 2015
Report Date
April 14, 2015
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE WAS MANUFACTURED IN JUNE 2014. THE DEVICE WAS RECEIVED AND THE EVALUATION IS COMPLETE. DURING THE EVENT HISTORY LOG REVIEW, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THE HOMECHOICE DEVICE RECEIVED A RETURNED INSTRUMENT TESTING EVALUATION (RITE).  THIS EVALUATION INCLUDED FUNCTIONAL AND ELECTRICAL TESTING OF THE DEVICE. A VISUAL INSPECTION WAS PERFORMED. A DEVICE HISTORY REVIEW REVEALED NO NONCONFORMITIES, REWORK, OR DEVIATIONS THAT WOULD CAUSE OR CONTRIBUTE TO THE ISSUE. THE REVIEW DID NOT IDENTIFY ANY ISSUES DURING DEVICE MANUFACTURING THAT ARE RELATED TO THE ISSUE OF IIPV. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THIS ISSUE WAS DETERMINED TO BE FALSE EMPTY DETECT AND USE ERROR, THE CLINICIAN INAPPROPRIATELY SET THE MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER'S GUIDE WARNS THAT "IF YOU SET THE MINIMUM DRAIN VOLUME % TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS EVENT OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2015 AT 12:31:53. DURING NIGHT DRAIN CYCLE SIX, THE PATIENT'S ULTRAFILTRATION READING WAS 1631ML, INDICATING THE HOME PATIENT DRAINED 1631ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2500ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302638 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 76 YR