FDA Adverse Event Injury Summary report: N

RESPIRONICS COMFORTCLASSIC MEDIUM CPAP MASK

MDR report key: 4758687 · Received May 5, 2015

Report

Report Number
MW5042567
Event Type
Injury
Date Received
May 5, 2015
Date of Event
April 25, 2015
Report Date
May 5, 2015
Manufacturer
RESPIRONICS
Product Code
NHJ
Adverse Event
Yes
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
IN, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

SUDDENLY DEVELOPED SEVERE SHORTNESS OF BREATH AND CHEST PAIN ON (B)(6) 2015 WHILE USING RESPIRONICS COMFORTCLASSIC MEDIUM CPAP MASK (PRODUCT NO. (B)(4), LOT NO. 140408). EVALUATED IN EMERGENCY ROOM (B)(6) 2015, AND FOUND TO HAVE RIGHT LUNG TENSION PNEUMOTHORAX ON CHEST X-RAY. IMMEDIATELY HAD RIGHT SIDED THORACOSTOMY WITH PLACEMENT OF CHEST TUBE UNDER VACUUM AND ADMITTED TO HOSPITAL. CHEST TUBE WAS REMOVED ON (B)(6) 2015 AND DISCHARGED FROM HOSPITAL ON (B)(6) 2015. AFTER DISCHARGE FROM HOSPITAL, CPAP MASK WAS CLOSELY EXAMINED AND WAS FOUND TO HAVE AN ALMOST COMPLETELY OCCLUDED EXHALATION PORT. THIS WAS CONFIRMED BY BEING UNABLE TO SLIP A PIECE OF PAPER INTO THE EXHALATION PORT. COMPARISON TO A DIFFERENT, OLD RESPIRONICS COMFORTCLASSIC MEDIUM CPAP MASK USING THE SAME PIECE OF PAPER SHOWED THAT THE OLD MASK HAD A PATENT EXHALATION PORT, UNLIKE THE MASK THAT IS SUSPECTED TO HAVE CAUSED THE PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294360 RESPIRONICS COMFORTCLASSIC MEDIUM CPAP MASK RESPIRONICS COMFORTCLASSIC MEDIUM CPAP MASK NHJ RESPIRONICS 1007964 140408

Patients

Seq Age Sex Outcome Treatment
1 59 YR Hospitalization| L| R| S RTC MEDS: CPAP AND TUBING WITH RESPIRONICS| COMFORTCLASSIC MEDIUM CPAP MASK.