FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4756805 · Received May 5, 2015

Report

Report Number
1052693-2015-00674
Event Type
Malfunction
Date Received
May 5, 2015
Date of Event
April 14, 2015
Report Date
December 21, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION BLACK CHEMISTRY OBSERVED, NO INVESTIGATION REQUIRED. INVESTIGATION COMPLETED AND METER EVALUATED WITH NO DEFECTS FOUND.

Additional Manufacturer Narrative · 1

PRODUCT NOT YET RETURNED FOR EVALUATION. (B)(4).

Description of Event or Problem · 1

CONSUMER COMPLAINT OF "LO" BLOOD RESULT. EXPECTED BLOOD GLUCOSE RESULTS FASTING RANGE FROM 95 TO 150 MG/DL. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. BACK TO BACK BLOOD TEST NO PERFORMED DUE TO EXPIRED TEST STRIPS. STORAGE OF TEST STRIPS NOT VERIFIED. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 06/30/2014 AND OPEN VIAL DATA NOT VERIFIED. RECALL TEST RESULTS PERFORMED NON-FASTING FROM METER MEMORY: "LO" (B)(6) 2015 07:37AM; 99 MG/DL (B)(6) 2014 09:24PM; 161 MG/DL (B)(6) 2014 10:55PM; 160MG/DL (B)(6) 2014 10:56PM; 166 MG/DL (B)(6) 2014 08:45PM.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF "LO" BLOOD RESULT. EXPECTED BLOOD GLUCOSE RESULTS FASTING RANGE FROM 95 TO 150MG/DL. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. BACK TO BACK BLOOD TEST NOT PERFORMED DUE TO EXPIRED TEST STRIPS. STORAGE OF TEST STRIPS NOT VERIFIED. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 06/30/2014 AND OPEN VIAL DATE NOT VERIFIED. RECALL TEST RESULTS PERFORMED NON-FASTING FROM METER MEMORY: "LO" (B)(6) 2015 07:37AM. 99MG/DL (B)(6) 2014 09:24PM . 161MG/DL (B)(6) 2014 10:55PM. 160MG/DL (B)(6) 2014 10:56PM. 166MG/DL (B)(6) 2014 08:45PM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
293437 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PP1346

Patients

Seq Age Sex Outcome Treatment
1 0 YR