FDA Adverse Event Death Summary report: N

AVALON FM20 FETAL MONITOR

MDR report key: 4756179 · Received May 7, 2015

Report

Report Number
9610816-2015-00093
Event Type
Death
Date Received
May 7, 2015
Date of Event
April 23, 2015
Report Date
April 29, 2015
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
HGM
PMA / PMN Number
K052795
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

PHILIPS MEDICAL SYSTEMS INVESTIGATED THE ISSUE DESCRIBED ABOVE. ON MAY 8, 2015, A PHILIPS FIELD SERVICE ENGINEER (FSE) PERFORMED A FUNCTIONAL TEST TO VERIFY THE FUNCTIONALITY OF THE AVALON FM20 FETAL MONITOR AND DETERMINED THAT THE DEVICE WORKED AS SPECIFIED. THE PROVIDED TRACE RECORDINGS WERE REVIEWED BY PHILIPS RESEARCH & DEVELOPMENT DEPARTMENT (R&D) AND SHOW: THE HEART RATE BASELINE WAS AROUND 120 BPM FROM 13:23 TO 13:38 AND FROM 14:16 TO 16:08 WITH AN INCREASING HEART RATE SIGNAL LOSS AFTER 15:00. MOVEMENTS WERE ALSO RECORDED. FROM 18:01 TO 18:48, ONLY A TOCO TRANSDUCER WAS APPLIED (TRACE ANNOTATION ¿FHRX MODE: NO TRANSDUCER¿). IN ADDITION A COPY OF THE FETAL MONITOR RECORDING WAS PROVIDED, COVERING ABOVE MENTIONED TIME FRAMES UNTIL 15:20. BASED ON THE REVIEW OF THE PROVIDED TRACES AND THE ADDITIONAL INFORMATION WE RECEIVED, IT WAS DETERMINED THAT THE PHILIPS DEVICE WORKED AS SPECIFIED. THERE IS NO INDICATION THAT THE CLINICIANS VERIFIED FETAL LIFE BEFORE INITIATING FETAL MONITORING DEVICE LABELING (INSTRUCTIONS FOR USE) INSTRUCTS THAT ¿FETAL MONITORING TECHNOLOGY AVAILABLE TODAY IS NOT ALWAYS ABLE TO DIFFERENTIATE A FETAL HEART RATE (FHR) SIGNAL SOURCE FROM A MATERNAL HEART RATE (MHR) SOURCE IN ALL SITUATIONS ¿. THIS PHENOMENON IS DUE TO LIMITATIONS OF THE TECHNOLOGY AND WIDELY INDEPENDENT OF THE BRAND AND MODEL OF THE FETAL MONITOR. IN ADDITION, AN INCREASED MATERNAL PULSE RATE AROUND AND ABOVE 120 BPM IS OFTEN SEEN WITH FETAL DEMISE. THE MATERNAL HEART RATE MAY BE ATYPICALLY HIGH AND THEREFORE CONFUSED WITH THAT OF A LIVE FETUS. APPARENT FETAL MOVEMENT (FMP) MAY ALSO BE DETECTED BY THE MONITOR BUT THIS MAY BE A RESULT OF MATERNAL MOVEMENT CAUSING THE FETUS TO MOVE WITHIN THE AMNIOTIC FLUID. PER PHILIPS (B)(4) AND FSE ENGINEER, THE M2702A AVALON FM20 AVALON FETAL MONITOR WITH S/N: (B)(4) WORKS AS SPECIFIED. NO PRODUCT MALFUNCTION. THE M2702A AVALON FM20 FETAL MONITOR REMAINS AT THE CUSTOMER SITE. THERE IS NO INDICATION OF ANY MALFUNCTION OF THE M2702A AVALON FM20 FETAL MONITOR. THE AVAILABLE INFORMATION DOES NOT SUPPORT THAT USE OF THE DEVICE WAS CONTRIBUTORY TO THE REPORTED STILLBIRTH. THE CAUSE OF THE STILLBIRTH IS UNKNOWN. BASED ON THE REVIEW OF THE PROVIDED TRACES AND THE ADDITIONAL INFORMATION WE RECEIVED, IT WAS DETERMINED THAT OUR DEVICE WORKED AS SPECIFIED. DEVICE LABELING (INSTRUCTIONS FOR USE) INSTRUCTS THAT ¿FETAL MONITORING TECHNOLOGY AVAILABLE TODAY IS NOT ALWAYS ABLE TO DIFFERENTIATE A FETAL HEART RATE (FHR) SIGNAL SOURCE FROM A MATERNAL HEART RATE (MHR) SOURCE IN ALL SITUATIONS ¿. USERS SHOULD CONFIRM FETAL LIFE BY INDEPENDENT MEANS BEFORE STARTING TO USE THE FETAL MONITOR. SINCE A MHR TRACE CAN EXHIBIT FEATURES THAT ARE VERY SIMILAR TO THOSE OF A FHR TRACE, USERS SHOULD NOT RELY SOLELY ON TRACE PATTERN FEATURES TO IDENTIFY A FETAL SOURCE. ALSO, FETAL MOVEMENT PROFILE (FMP) ANNOTATIONS ON A FETAL TRACE ALONE MAY NOT ALWAYS INDICATE THAT THE FETUS IS ALIVE. THE BODY OF A DECEASED FETUS CAN MOVE AND CAUSE THE MONITOR TO ANNOTATE. MATERNAL SIGNAL SOURCES MAY BE PICKED UP WHEN USING THE ULTRASOUND TRANSDUCER AND COULD LEAD TO MISIDENTIFICATION WHEN THE MHR IS HIGHER THAN NORMAL (ESPECIALLY WHEN > 100 BPM).

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE SUBMITTED AFTER PHILIPS OBTAINS MORE INFORMATION CONCERNING THIS EVENT.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THEY WERE "MONITORING WITH M2702A AVALON FM20 FETAL MONITOR S/N:(B)(4). BABY WAS BORN DEAD BUT THE FETAL MONITOR HAS RECORDED FETAL HEART RATE (FHR) TRACES AND FETAL MOVEMENTS (FMP) ALTHOUGH THE FETUS ALREADY WAS DEAD. USER WANTS TO KNOW HOW TO IDENTIFY IT WAS MOTHER¿S HR INSTEAD OF FHR. THERE WAS FMP AND CONTRACTION SHOWED IN THE TOCO."

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THEY WERE "MONITORING WITH M2702A AVALON FM20 FETAL MONITOR S/N: (B)(4). BABY WAS BORN DEAD BUT THE FETAL MONITOR HAS RECORDED FETAL HEART RATE (FHR) TRACES AND FETAL MOVEMENTS (FMP) ALTHOUGH THE FETUS ALREADY WAS DEAD. USER WANT TO KNOW HOW TO IDENTIFY IT WAS MOTHER'S HR INSTEAD OF FHR. THERE WAS FMP AND CONTRACTION SHOWED IN THE TOCO".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300443 AVALON FM20 FETAL MONITOR FETAL MONITOR HGM PHILIPS MEDICAL SYSTEMS M2702A

Patients

Seq Age Sex Outcome Treatment
1 Death