FDA Adverse Event Malfunction Summary report: N

DEPTH GAUGE FOR 2.7MM & SMALL SCREWS

MDR report key: 4745963 · Received May 4, 2015

Report

Report Number
1719045-2015-10270
Event Type
Malfunction
Date Received
May 4, 2015
Report Date
April 14, 2015
Manufacturer
SYNTHES HAGENDORF
Product Code
HTJ
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SERVICE HISTORY REVIEW: LOT #8935221- NO SERVICE HISTORY REVIEW CAN BE PERFORMED AS THIS IS A LOT CONTROLLED ITEM. THE MANUFACTURE DATE OF THIS ITEM IS JULY 16, 2014. THE SOURCE OF THE MANUFACTURE DATE IS THE RELEASE TO WAREHOUSE DATE. PRODUCT INVESTIGATION SUMMARY: ONE OF THE FOLLOWING WAS RECEIVED: DEPTH GAUGE (PART 319.04 / LOT 8935221). THE RETURNED DEPTH GAUGE RECEIVED ASSEMBLED WAS COMPLETELY DISASSEMBLED. THE NIPPLE/COLLAR IS MISSING AND WAS NOT RETURNED, AND THE HOOKED NEEDLE HAS BEEN BROKEN OFF THE SLIDER BODY. THE DEVICE DOES NOT APPEAR TO HAVE ANY FURTHER DAMAGE. A VISUAL INSPECTION AND DRAWING REVIEW WERE PERFORMED AS PART OF THIS INVESTIGATION. NO PRODUCT DESIGN ISSUES OR DISCREPANCIES WERE OBSERVED. THE CAUSE OF THIS COMPLAINT CONDITION CANNOT BE DETERMINED, BUT IT WAS MOST LIKELY CAUSED BY INATTENTIVENESS DURING STERILIZATION. THIS COMPLAINT IS CONFIRMED. THE ASSOCIATED DRAWING FOR THE DEVICE WAS REVIEWED. RELEVANT FEATURES/DIMENSIONS WERE INSPECTED AND FOUND TO BE WITHIN SPECIFICATION. NO DRAWING ISSUES OR DISCREPANCIES WERE NOTED. THE DESIGN IS ADEQUATE FOR ITS INTENDED USE AND DID NOT CONTRIBUTE TO THIS COMPLAINT CONDITION. NO OTHER ISSUES WERE IDENTIFIED WITH THE DEVICE. THE DESIGN IS ADEQUATE FOR ITS INTENDED USE, AND IT DID NOT CONTRIBUTE TO THE COMPLAINT CONDITION. DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS.IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. EVENT DATE: UNKNOWN. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. A SERVICE HISTORY RECORD REVIEW HAS BEEN REQUESTED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

(B)(6) REPORTED THE FOLLOWING EVENT: IT WAS REPORTED THAT THE MEASURING NEEDLE IS BROKEN AND THE COLLAR IS MISSING FROM THE DEPTH GAUGE. THERE WAS NO REPORT OF PATIENT OR SURGICAL INVOLVEMENT ASSOCIATED WITH THE COMPLAINT EVENT. THIS REPORT IS 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291022 DEPTH GAUGE FOR 2.7MM & SMALL SCREWS GAUGE,DEPTH HTJ SYNTHES HAGENDORF 8935221

Patients

Seq Age Sex Outcome Treatment
1