LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø10X300MM X 125°
Report
- Report Number
- 0009610622-2015-00223
- Event Type
- Injury
- Date Received
- May 4, 2015
- Date of Event
- March 13, 2015
- Report Date
- April 16, 2015
- Manufacturer
- STRYKER TRAUMA KIEL
- Product Code
- HSB
- PMA / PMN Number
- K034002
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION SUMMARY: EVALUATION REVEALED THE LONG GAMMA NAIL TO BE THE PRIMARY PRODUCT. ALL OTHER MENTIONED PRODUCTS ARE REGARDED TO BE CONCOMITANT ITEMS. A REVIEW OF THE DHR REVEALED NO DISCREPANCIES IN MATERIAL OR MANUFACTURING. DISCUSSION OF THE DAMAGES OBSERVED ON PRIMARY PRODUCT. NAIL: DAMAGE AND EVIDENCE OF THE BREAKAGE SURFACES, COURSE OF THE BREAKAGE LINE AS WELL AS THE ABSENCE OF PLASTIC DEFORMATION (ALONG THE BREAKAGE LINE) SUGGEST THAT THE IMPLANT FRACTURED IN A FATIGUE MANNER DUE TO OVERLOAD. EVIDENT LINES OF REST FOUND ON BOTH FRAGMENTS INDICATE A FATIGUE FRACTURE OF THE NAIL RUNNING FROM LATERAL TO MEDIAL. APPEARANCE OF THE BREAKAGE SURFACES, AS WELL AS THE COURSE OF THE BREAKAGE LINE, WORN AREAS AND MATERIAL DISPLACEMENT IN OUTSIDE DIRECTION SUGGESTS THAT MOST LIKELY THE FATIGUE FRACTURE STARTED IN THE ANTERIOR WEB AND WITH A CERTAIN DELAY IN THE POSTERIOR WEB AS WELL. BENT OUTWARD MATERIAL AND STRUCTURE OF THE BREAKAGE SURFACE SUGGESTS THE RESIDUAL (FORCED-FRACTURE) BEING LOCATED AT THE MEDIAL TIP OF THE POSTERIOR WEB. EVIDENT MATERIAL ABRASION AND DEFORMATION (FRICTION MARKS IN THE SCREW HOLE) WERE IDENTIFIED AS COLD SETS (FRETTING CORROSION), WHICH WERE CAUSED BY HIGH LOAD BEARING WITH THE LAG SCREW DURING THE PERIOD OF IMPLANTATION. THE LATERAL EDGE OF THE ANTERIOR WEB OF THE PROXIMAL DRILL HOLE HAD BECOME SIGNIFICANTLY DAMAGED DUE TO CONTACT WITH THE LAG SCREW STEPDRILL. SUCH ¿ NOT INTENDED ¿ MATERIAL DAMAGES CAN CAUSE ADDITIONAL NOTCH EFFECTS. DISTAL DRILL HOLES: DAMAGE OBSERVED WAS CAUSED DURING SCREW INSERTION/REMOVAL AND WORN AREAS INDICATE HIGH/PERMANENT LOAD BEING APPLIED TO THE NAIL. LAG SCREW OF U-BLADE SET: THE DAMAGE OF THE SHAFT WAS IDENTIFIED AS COLD SETS, WHICH WERE CAUSED DUE TO HIGH RESPECTIVELY PERMANENT LOAD BEARING. THE MARK AT THE EDGE OF THE FLUTE INDICATES THAT THE SET SCREW HAS BEEN INSERTED AS INTENDED TO ENSURE LOCKING OF THE LAG SCREW AGAINST ROTATION AND MIGRATION. LOCKING SCREW Ø5X37,5MM (K0F045A): DAMAGE OBSERVED INDICATES HIGH/PERMANENT LOAD APPLICATION. LOCKING SCREW Ø5X37,5MM (K0E5BA1): DAMAGE AND EVIDENCE OF THE BREAKAGE SURFACE, COURSE OF THE BREAKAGE LINE AS WELL AS THE ABSENCE OF PLASTIC DEFORMATION (ALONG THE BREAKAGE LINE) SUGGEST THAT THE IMPLANT FRACTURED IN A FATIGUE MANNER DUE TO OVERLOAD. EVIDENT LINES OF REST FOUND ON THE BREAKAGE SURFACE INDICATE A FATIGUE FRACTURE OF THE SCREW STARTING AT THE THREAD MINOR DIAMETER AND PROCEEDING THROUGH THE ENTIRE CROSS SECTION. DAMAGE OBSERVED INDICATES HIGH/PERMANENT LOAD APPLICATION. GENERAL TECHNICAL ASPECTS: THE BROKEN IMPLANTS ARE TEMPORARY IMPLANTS WHICH INEVITABLY WILL BE SUBJECT OF A FATIGUE FRACTURE IF THE STRESSES ON THE IMPLANT ARE TOO HIGH OR NOT CONSIDERABLY REDUCED DURING THE PERIOD OF IMPLANTATION. THIS COULD BE DESCRIBED AS A RACE BETWEEN BONY CONSOLIDATION / FRACTURE HEALING AND IMPLANT BREAKAGE. IF FRACTURE HEALING IS INSUFFICIENT MECHANICAL DAMAGE OF A LOAD-BEARING OR LOAD-SHARING DEVICE IS MORE LIKELY. GENERALLY THE RISK OF A BREAKAGE WILL INCREASE WITH THE INCREASE OF LOAD CYCLES AND LOAD LEVEL. FINAL CONCLUSION: THE NAIL BROKE AFTER AN IMPLANTATION PERIOD OF APPROXIMATELY 11 MONTHS. INFORMATION ABOUT BONE HEALING WAS NOT PROVIDED. NEVERTHELESS WITH REGARD TO THE LONG IMPLANTATION PERIOD IT CAN BE CONCLUDED THAT THE LEVEL OF STRESS, WHICH DID NOT REDUCE DURING THE PERIOD OF IMPLANTATION CONTRIBUTED SIGNIFICANTLY TO THE BREAKAGE OF THE IMPLANT(S). ADDITIONALLY AN INTRAOPERATIVE DAMAGE OF THE GAMMA NAIL CAUSED BY A DEVIATED STEP DRILL RESULTING IN SIGNIFICANT WEAKENING OF THE GAMMA NAIL IN THE AREA OF THE LAG SCREW THRU HOLE.
ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2014, THE PATIENT UNDERWENT THE SURGERY WITH THE G3 LONG NAIL. ON (B)(6) 2015, THE PATIENT FELL, AND NAIL BROKE. THEREFORE, THE PATIENT UNDERWENT THE REVISION SURGERY ON (B)(6) 2015.
ON (B)(6) 2014, THE PATIENT UNDERWENT THE SURGERY WITH THE G3 LONG NAIL. ON (B)(6) 2015, THE PATIENT FELL, AND NAIL BROKE. THEREFORE, THE PATIENT UNDERWENT THE REVISION SURGERY ON (B)(6) 2015.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 291434 | LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø10X300MM X 125° | ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES | HSB | STRYKER TRAUMA KIEL | K0DC0DE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention |