FDA Adverse Event Malfunction Summary report: N

CELL-DYN 3700 SL ANALYZER

MDR report key: 4743471 · Received May 1, 2015

Report

Report Number
2919069-2015-00024
Event Type
Malfunction
Date Received
May 1, 2015
Date of Event
March 25, 2015
Report Date
March 27, 2015
Manufacturer
ABBOTT DIAGNOSTICS DIVISION
Product Code
GKZ
PMA / PMN Number
K980614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. (B)(4). AN EVALUATION IS IN PROCESS.

Additional Manufacturer Narrative · 1

FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, A REVIEW OF LABELING AND DATA LOG REVIEWS. NO ADVERSE TREND WAS IDENTIFIED FOR THE CUSTOMER'S ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. CELL-DYN 3700, LIST NUMBER 02H31-01, SN (B)(4), TRANSMITTING ONE PATIENT RESULT EIGHT TIMES WITH EIGHT DIFFERENT PATIENT IDENTIFICATION NUMBERS. THE 8 PATIENT RESULTS SUBMITTED SHOWED THE SAMPLES WERE RUN ON (B)(6) 2015 BETWEEN 12:53 TO 14:26, AND WERE IDENTIFIED AS (B)(6). THE DATALOG SHOWED THAT CORRECT RESULTS WERE GENERATED FOR EACH OF THESE 8 PATIENT SAMPLES. THE RESULT DATA FOR (B)(6), WAS RUN ON (B)(6) 2015 AT 14:28. THE RESULT OF (B)(6) WAS TRANSMITTED IN THE LABORATORY INFORMATION SYSTEM (LIS) 8 TIMES BUT WAS ASSIGNED TO 8 DIFFERENT PATIENT IDENTIFICATION NUMBERS. THE CELL-DYN 3700 TRANSMIT LOG WAS NOT SUBMITTED BECAUSE IT WAS NOT AVAILABLE. THE ACTUAL DATA SENT BY THE CELL-DYN 3700 INSTRUMENT DATA STATION COULD NOT BE VERIFIED. DUE TO THE MISSING TRANSMIT LOG DATA FROM THE CELL-DYN 3700 DATA STATION, THE EXACT CAUSE OF THIS INCIDENT WAS UNDETERMINED BECAUSE THE COMMUNICATION WAS UNABLE TO BE VERIFIED BETWEEN THE CELL-DYN 3700 AND THE LIS. BASED ON ALL AVAILABLE INFORMATION AND ABBOTT DIAGNOSTICS' COMPLAINT INVESTIGATION NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

THE CUSTOMER OBSERVED THE CELL-DYN 3700 ANALYZER TRANSMITTING ONE PATIENT RESULT EIGHT TIMES WITH EIGHT DIFFERENT PATIENT IDENTIFICATION NUMBERS. THE 8 PATIENT RESULTS SUBMITTED FROM CELL-DYN 3700 SL, LIST NUMBER 02H31-01, SERIAL NUMBER (B)(4), SHOWED THE SAMPLES WERE RUN ON (B)(6) 2015 BETWEEN 12:53 TO 14:26, AND WERE IDENTIFIED AS SEQ. 1295 TO SEQ. 1349. THE DATALOG SHOWED THAT CORRECT RESULTS WERE GENERATED FOR EACH OF THESE 8 PATIENT SAMPLES. THE RESULT DATA FOR SEQ. 1351, SPECIMEN ID (B)(6), WAS RUN ON (B)(6) 2015 AT 14:28. THE RESULT OF SEQ.1351 WAS TRANSMITTED IN THE LABORATORY INFORMATION SYSTEM (LIS) 8 TIMES BUT WAS ASSIGNED TO 8 DIFFERENT PATIENT IDENTIFICATION NUMBERS. THE DATA FROM THE LIS SHOWED THE COMPLAINT INCIDENT; HOWEVER, THE CELL-DYN 3700 TRANSMIT LOG WAS NOT SUBMITTED BECAUSE IT WAS NOT AVAILABLE. THE ACTUAL DATA SENT BY THE CELL-DYN 3700 INSTRUMENT DATA STATION COULD NOT BE VERIFIED. DUE TO THE MISSING TRANSMIT LOG DATA FROM THE CELL-DYN 3700 DATA STATION, THE COMMUNICATION BETWEEN THE CELL-DYN 3700 AND THE LIS COULD NOT BE VERIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
288056 CELL-DYN 3700 SL ANALYZER AUTOMATED HEMATOLOGY ANALYZER GKZ ABBOTT DIAGNOSTICS DIVISION

Patients

Seq Age Sex Outcome Treatment
1