FDA Adverse Event
Other
Summary report: N
MCIVOR MOUTH GAG FRAME
MDR report key: 474110
·
Received July 28, 2003
Report
- Report Number
- 474110
- Event Type
- Other
- Date Received
- July 28, 2003
- Date of Event
- May 1, 2003
- Report Date
- June 1, 2003
- Manufacturer
- PILLING SURGICAL
- Product Code
- KBN
- Report Source
- User Facility report
- Reporter Location
- NC, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
PT WITH OBSTRUCTIVE PALATINE TONSILS UNDERWENT A TONSILLECTOMY. A MCIVOR MOUTH GAG WAS PLACED, AND THE RIGHT TONSIL WAS REMOVED WITHOUT INCIDENT. BLEEDING WAS CONTROLLED WITH CAUTERY. THE IDENTICAL PROCEDURE WAS CARRIED OUT ON THE LEFT SIDE, AND ACCORDING TO THE OPERATIVE NOTE, A BURN WAS NOTICED TO THE LEFT ANTERIOR TONGUE THAT APPEARED TO HAVE OCCURRED BY TRANSDUCTION OF ELECTRICAL CURRENT THROUGH THE MOUTH GAG. THE BURNED AREA WAS APPROX 1 X 2CM IN DIMENSION BUT WAS VERY SUPERFICIAL. THE NON-VIABLE TISSUE WAS EXCISED AND THE WOUND WAS CLOSED WITH SUTURES. THE SURGEON BELIEVES THAT BETTER DESIGN OF THE INSTRUMENTS SO THAT THEY ARE INSULATED WOULD HAVE PREVENTED THIS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MCIVOR MOUTH GAG FRAME | MOUTH GAG FRAME | KBN | PILLING SURGICAL | * | * | |
| 2 | BLADES | OTHER | KBN | PILLING SURGICAL | * | * | |
| 3 | SABRE 2400 | BIPOLAR ESU | GEI | CONMED | SABRE 2400 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Other | ELECTROCAUTERY. |