FDA Adverse Event Other Summary report: N

MCIVOR MOUTH GAG FRAME

MDR report key: 474110 · Received July 28, 2003

Report

Report Number
474110
Event Type
Other
Date Received
July 28, 2003
Date of Event
May 1, 2003
Report Date
June 1, 2003
Manufacturer
PILLING SURGICAL
Product Code
KBN
Report Source
User Facility report
Reporter Location
NC, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

PT WITH OBSTRUCTIVE PALATINE TONSILS UNDERWENT A TONSILLECTOMY. A MCIVOR MOUTH GAG WAS PLACED, AND THE RIGHT TONSIL WAS REMOVED WITHOUT INCIDENT. BLEEDING WAS CONTROLLED WITH CAUTERY. THE IDENTICAL PROCEDURE WAS CARRIED OUT ON THE LEFT SIDE, AND ACCORDING TO THE OPERATIVE NOTE, A BURN WAS NOTICED TO THE LEFT ANTERIOR TONGUE THAT APPEARED TO HAVE OCCURRED BY TRANSDUCTION OF ELECTRICAL CURRENT THROUGH THE MOUTH GAG. THE BURNED AREA WAS APPROX 1 X 2CM IN DIMENSION BUT WAS VERY SUPERFICIAL. THE NON-VIABLE TISSUE WAS EXCISED AND THE WOUND WAS CLOSED WITH SUTURES. THE SURGEON BELIEVES THAT BETTER DESIGN OF THE INSTRUMENTS SO THAT THEY ARE INSULATED WOULD HAVE PREVENTED THIS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MCIVOR MOUTH GAG FRAME MOUTH GAG FRAME KBN PILLING SURGICAL * *
2 BLADES OTHER KBN PILLING SURGICAL * *
3 SABRE 2400 BIPOLAR ESU GEI CONMED SABRE 2400 *

Patients

Seq Age Sex Outcome Treatment
1 25 YR Other ELECTROCAUTERY.