ANSPACH EMAX 2 PLUS BURR MOTOR
Report
- Report Number
- 3005985723-2015-00058
- Event Type
- Malfunction
- Date Received
- April 30, 2015
- Date of Event
- January 16, 2015
- Report Date
- March 27, 2015
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K112507
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED AT MAKO SURGICAL. EVALUATION OF THE RETURNED MOTORS IS ONGOING. A SUPPLEMENTAL REPORT WILL BE FILED IF ADDITIONAL RESULTS ARE OBTAINED.
REPORTED EVENT: EVENT REPORTED ANSPACH MOTOR OVERHEATING AND SHUTTING OFF DURING SURGERY. THERE WAS NO HARM TO PATIENT OR USER. THE REPORTED EVENT WAS CONFIRMED. METHOD & RESULTS: DEVICE EVALUATION AND RESULTS: DEVICE EVALUATION WAS PERFORMED AND THE ANSPACH EMAX 2 PLUS BURR MOTOR EVENT WAS CONFIRMED. DEVICE HISTORY REVIEW: NOT PERFORMED AS THE ANSPACH EMAX 2 PLUS BURR MOTOR IS AN OEM PRODUCT. COMPLAINT HISTORY REVIEW: BASED ON THE DEVICE IDENTIFICATION, THE CATSWEB AND TRACKWISE COMPLAINT DATABASES WERE REVIEWED FROM 2011 TO PRESENT FOR SIMILAR REPORTED EVENTS REGARDING OVERHEATING OF P/N 110940 S/N (B)(4). CONCLUSIONS: THE ANSPACH MOTOR IS AN OEM DEVICE. UPON RECEIPT, THE DEVICE WAS EVALUATED PER (B)(4) ANSPACH EMAX 2 PLUS BURR MOTOR AND THE REPORTED EVENT WAS CONFIRMED.
THE SURGEON WAS PERFORMING A PARTIAL KNEE ARTHROPLASTY USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS MULTICOMPARTMENTAL KNEE (MCK) IMPLANTS. DURING THE CASE, THE BURR MOTOR OVERHEATED AND SHUTDOWN. THE BACKUP BURR MOTOR WAS ALSO UNAVAILABLE DURING THE SAME CASE.
THE SURGEON WAS PERFORMING A PARTIAL KNEE ARTHROPLASTY USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS MULTICOMPARTMENTAL KNEE (MCK) IMPLANTS. DURING THE CASE, THE BURR MOTOR OVERHEATED AND SHUTDOWN. THE BACKUP BURR MOTOR WAS ALSO UNAVAILABLE DURING THE SAME CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 285732 | ANSPACH EMAX 2 PLUS BURR MOTOR | STEROTAXIC DEVICE, ACCESSORY | OLO | MAKO SURGICAL CORP. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |