FDA Adverse Event Malfunction Summary report: N

IN SITU BENDER-RIGHT

MDR report key: 4740989 · Received April 30, 2015

Report

Report Number
9680938-2015-10039
Event Type
Malfunction
Date Received
April 30, 2015
Date of Event
April 14, 2015
Report Date
April 14, 2015
Manufacturer
SYNTHES TUTTLINGEN
Product Code
HXP
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. NO PATIENT INFORMATION REPORTED. ADDITIONAL PRODUCT CODE: HWX. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ONGOING; NO CONCLUSION COULD BE DRAWN AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. DEVICE HISTORY RECORDS WAS CONDUCTED. THE REPORT INDICATES THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE RAW MATERIAL WHICH WAS DELIVERED AS LOT #00057650 IS CORRESPONDING TO THE SPECIFICATIONS. THE HARDNESS WAS MEASURED AT THE TIME OF THE MANUFACTURING AT 46,9 HRC AND WAS FOUND TO BE GOOD. NO NCRS WERE GENERATED DURING PRODUCTION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. A PRODUCT INVESTIGATION WAS COMPLETED: ONE IN SITU ROD BENDER (388.112, LOT T969971) WAS RETURNED WITH A COMPLAINT CONDITION OF A BROKEN JAW ON THE STRAIGHT END OF THE INSTRUMENT. THE COMPLAINT CONDITION WAS CONFIRMED AFTER EXAMINATION AS AN 11MM FRAGMENT BROKE OFF THE JAW. A DEFINITIVE ROOT CAUSE WAS UNABLE TO BE DETERMINED; HOWEVER THE FAILURE IS CONSISTENT WITH THE APPLICATION OF EXCESSIVE FORCE TO THE INSTRUMENT (MANUFACTURED SEPTEMBER 2011). REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE RAW MATERIAL IS CORRESPONDING TO THE SPECIFICATIONS. THE HARDNESS WAS MEASURED AT THE TIME OF THE MANUFACTURING AND WAS FOUND TO BE GOOD. NO NON-CONFORMANCE REPORTS WERE GENERATED DURING PRODUCTION. THE RETURNED INSTRUMENT WAS EXAMINED AND THE COMPLAINT CONDITION WAS ABLE TO BE CONFIRMED. THE MATERIAL AT THE FRACTURE SITE APPEARS TO BE HOMOGENOUS WITHOUT SIGNS OF FATIGUE. THE FAILURE MODE IS CONSISTENT WITH A MECHANICAL OVERLOAD. THE COMPLAINT IS CONFIRMED. PER THE TECHNIQUE GUIDES, THE BENDERS ARE INCLUDED IN USS, PANGEA, AND CLICK¿X SYSTEMS FOR IN SITU SAGITTAL PLANE BENDING OF RODS. ALTERNATIVELY, ROD BENDERS 388.91, 388.92, 388.112, 388.113, 388.117, 388.961, 03.620.020, 03.622.060, 03.622.061, AND 03.622.062 MAY BE USED TO ACHIEVE CONTOURING OF THE RODS. RELEVANT DRAWINGS FOR THE RETURNED INSTRUMENT WERE REVIEWED, BOTH CURRENT AND FROM THE TIME OF MANUFACTURING. THE BENDER IS MADE FROM H900 HEAT TREATED 17-4PH AND IS APPROPRIATE FOR THE INTENDED USE OF THE INSTRUMENTS AS H900 PROVIDES MAXIMUM HARDNESS. WEAR COUPLED WITH USER TECHNIQUE MAY HAVE CAUSED THE COMPLAINT CONDITION HOWEVER THE USER TECHNIQUE IS UNKNOWN AND SO THE ROOT CAUSE IS INDETERMINATE. THE DESIGN, MATERIALS AND FINISHING PROCESSES WERE FOUND TO BE APPROPRIATE FOR THE INTENDED USE OF THESE DEVICES. A DEFINITIVE ROOT CAUSE WAS UNABLE TO BE DETERMINED; HOWEVER THE FAILURE IS CONSISTENT WITH THE APPLICATION OF EXCESSIVE FORCE TO THE INSTRUMENT. THE CALCULATED OCCURRENCE RATE IS ACCEPTABLE UNDER THE SYSTEM RISK ASSESSMENT. ADDITIONALLY NO DESIGN DEFICIENCIES WERE IDENTIFIED WHICH WOULD CONTRIBUTE TO THE COMPLAINT CONDITION. PATIENT WAS INVOLVED BUT PATIENT ID IS UNKNOWN DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN INITIAL T2-L1 POSTERIOR FUSION SURGERY THE IN-SITU ROD BENDER-RIGHT BROKE. IT WAS REPORTED THE PATIENT HAD SCOLIOSIS AND HAD HARD BONES. AFTER PLACING THE ROD AND LOADING ALL THE CAPS, THE SURGEON BEGAN USING THE ROD BENDER, HE BENT THE ROD IN A COUPLE OF AREAS BUT THE BENDER SNAPPED ON THE LAST AREA THAT HE WANTED TO BEND. A SINGLE BROKEN SHARD WAS GENERATED THAT WAS SUCCESSFULLY RETRIEVED FROM THE PATIENT AND PLACED ALONGSIDE THE INSTRUMENT AT THE BACK TABLE. THE SURGERY CONTINUED AND THE SURGEON WAS HAPPY WITH HOW THE ROD LOOKED DESPITE THE BROKEN INSTRUMENT. THERE WAS NO HARM TO THE PATIENT AND SURGERY WAS DELAYED BY TWO TO THREE MINUTES. THIS REPORT IS 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
285812 IN SITU BENDER-RIGHT INSTRUMENT, BENDING OR CONTOURING HXP SYNTHES TUTTLINGEN T969971

Patients

Seq Age Sex Outcome Treatment
1