FDA Adverse Event Other Summary report: N

VISUAL PACS

MDR report key: 473903 · Received July 22, 2003

Report

Report Number
2135350-2003-00002
Event Type
Other
Date Received
July 22, 2003
Date of Event
June 26, 2003
Report Date
July 22, 2003
Manufacturer
EMAGEON UV, INC.
Product Code
LLZ
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

CLICKING ON ONE STUDY CAN LAUNCH ANOTHER FOR THE SAME PT: A STUDY IS LAUNCHED. THE STUDY IS DISPLAYED AS A PRIOR, AND ANOTHER STUDY IS SHOWN INSTEAD FOR THE SAME PT. THIS OCCURS WHEN THE STUDY TIME OF THE SELECTED STUDY IS BLANK. THE LIST OF PRIORS IS DETERMINED AS ALL STUDIES THAT FALL ON THAT DAY OR BEFORE. HOWEVER, THE STUDIES ARE THEN SORTED ASSUMING THE CURRENT STUDY'S TIME IS 00:00:00 MAKING ANY STUDY WITH A NEWER STUDY DATE SHOW UP AS THE CURRENT STUDY. THIS APPEARS TO BE HAPPENING WHEN TWO STUDIES EXIST FOR THE SAME DAY AND ONE STUDY HAS A NULL STUDY TIME, SERIES DATE AND SERIES TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VISUAL PACS MEDICAL IMAGE MANAGEMENT DEVICE LLZ EMAGEON UV, INC. 2.2.1D NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other