FDA Adverse Event
Other
Summary report: N
VISUAL PACS
MDR report key: 473903
·
Received July 22, 2003
Report
- Report Number
- 2135350-2003-00002
- Event Type
- Other
- Date Received
- July 22, 2003
- Date of Event
- June 26, 2003
- Report Date
- July 22, 2003
- Manufacturer
- EMAGEON UV, INC.
- Product Code
- LLZ
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
CLICKING ON ONE STUDY CAN LAUNCH ANOTHER FOR THE SAME PT: A STUDY IS LAUNCHED. THE STUDY IS DISPLAYED AS A PRIOR, AND ANOTHER STUDY IS SHOWN INSTEAD FOR THE SAME PT. THIS OCCURS WHEN THE STUDY TIME OF THE SELECTED STUDY IS BLANK. THE LIST OF PRIORS IS DETERMINED AS ALL STUDIES THAT FALL ON THAT DAY OR BEFORE. HOWEVER, THE STUDIES ARE THEN SORTED ASSUMING THE CURRENT STUDY'S TIME IS 00:00:00 MAKING ANY STUDY WITH A NEWER STUDY DATE SHOW UP AS THE CURRENT STUDY. THIS APPEARS TO BE HAPPENING WHEN TWO STUDIES EXIST FOR THE SAME DAY AND ONE STUDY HAS A NULL STUDY TIME, SERIES DATE AND SERIES TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VISUAL PACS | MEDICAL IMAGE MANAGEMENT DEVICE | LLZ | EMAGEON UV, INC. | 2.2.1D | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |