MICROCLAVE CONNECTOR
Report
- Report Number
- 2025816-2015-00022
- Event Type
- Malfunction
- Date Received
- April 28, 2015
- Date of Event
- February 16, 2015
- Report Date
- February 20, 2015
- Manufacturer
- ICU MEDICAL, INC.
- Product Code
- FPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
DEVICE RETURN: THREE (3) USED 011-C3300 MICROCLAVE CONNECTORS AND ONE (1) PACKAGED 011-C3300, LOT 2930437; USED ARIES ADMIN EXT SET; ARIES STOPCOCK; BD VENFLON PRO CATHETER AND BD 10ML SYRINGE. VISUAL INSPECTION RECORDED TWO OF THE USED MICROCLAVE CONNECTORS SILICONE PLUGS WERE PROTRUDED; NO OTHER VISUAL ABNORMALITIES NOTED WITH THE REMAINING SAMPLES THAT WERE RETURNED. FUNCTIONAL TESTING: THE USED 011-C3300 AND PKGD SAME LOT SAMPLE WERE EACH PRESSURE LEAK TESTED PER THE APPLICABLE PROD SPEC. THE RESULTS RECORDED NO FUNCTIONAL ISSUES, THE SILICONE PLUGS RESET CORRECTLY, THERE WERE NO LEAKAGES AND OR OUT OF SPEC CONDITIONS. THE RETURNED MATING AND ACCESS DEVICES WERE CONNECTED TO REPLICATE THE USAGE SET-UPS AND TESTED. THE REPORT NOTED THE ONLY FEMALE LUER PORT AVAILABLE TO CONNECT A 011-C3300 MICROCLAVE WAS THE SIDEPORT OF THE STOPCOCK HANDLE, WHICH WAS TURNED OFF TO THE SIDEPORT. THE RESULTS SHOWED NO FLOW OR PERFORMANCE ISSUES, THE 011-C3300 CONNECTOR DEVICE/COMPONENTS PERFORMED AS EXPECTED.
INTL ((B)(6)) COMPLAINT REC'D REPORTING COMPONENT/PLUG PROTRUSIONS WITH USE OF TWO 011-C3300 MICROCLAVE CONNECTORS. LIMITED INFORMATION REC'D, DISTRIBUTOR REPORTS "SOMETIMES WHEN THEY DO THE WASHING CAP FORGET TO CLOSE TO THE CONNECTOR". THERE WERE NO REPORTED ADVERSE PT CONSEQUENCES. ADD'L RELEVANT EVENT INFO ALTHOUGH REQUESTED HAS NOT BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279681 | MICROCLAVE CONNECTOR | MICROCLAVE CONNECTOR | FPA | ICU MEDICAL, INC. | 011-C3300 | 2930437 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | ARIES STOPCOCK| BD VENFLON PRO CATHETER| ARIES ADMIN EXT SET| BD 10ML SYRINGE |