FDA Adverse Event Injury Summary report: N

GEBAUER'S ETHYL CHLORIDE MIST SPRAY

MDR report key: 4736168 · Received April 28, 2015

Report

Report Number
1519179-2015-00001
Event Type
Injury
Date Received
April 28, 2015
Date of Event
March 27, 2015
Report Date
April 27, 2015
Manufacturer
GEBAUER CO.
Product Code
MLY
PMA / PMN Number
K991514
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT AS REPORTED IDENTIFIED THAT THE PT FELT A BURNING SENSATION IMMEDIATELY AFTER THE PRODUCT WAS APPLIED THAT RESULTED IN BLISTERING, SCABBING AND DISCOLORATION. THESE ADVERSE EVENTS ARE TYPICALLY ASSOCIATED WITH FROSTBITE THAT COULD RESULT FROM OVER-APPLICATION OF THE DEVICE. THE COMPLAINANT, WHO DID NOT SPEAK TO THE PT DIRECTLY WAS TOLD THE PRODUCT WAS APPLIED BY A NURSE FOR APPROX 5 SECONDS. A 5 SECOND SPRAY IS WITHIN THE IFU BUT IF THE SKIN HAD TURNED WHITE PRIOR TO THE 5 SECONDS, THE SPRAY SHOULD HAVE BEEN STOPPED PER THE DIRECTIONS FOR USE. ALSO, ALTHOUGH RARE, THE ADVERSE EVENT COULD BE THE RESULT OF AN ALLERGIC REACTION TO ETHYL CHLORIDE OR URTICARIA. GEBAUER WAS TOLD THE PT REQUIRED MEDICAL ATTENTION BUT NO INFO WAS PROVIDED AS TO WAT INTERVENTIONS WERE TAKEN. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN THE UNIT BACK AS WELL AS SEEK FURTHER INFO ON THE INJURY AND THE MEDICAL ATTENTION REQUIRED BUT WE WERE INFORMED THE PRODUCT WOULD NOT BE RETURNED AND NO FURTHER INFO WAS AVAILABLE.

Description of Event or Problem · 1

THE PT WAS RECEIVING DIALYSIS AND GEBAUER'S ETHYL CHLORIDE MIST SPRAY (LOT 3068) WAS APPLIED FOR PRE-INJECTION ANESTHESIA. THE PT REPORTED BURNING, BLISTERING, SCABBING AND DISCOLORATION AFTER APPLICATION OF THE PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
279053 GEBAUER'S ETHYL CHLORIDE MIST SPRAY CAPOCOOLANT MLY GEBAUER CO. 0386-0001-02 3068

Patients

Seq Age Sex Outcome Treatment
1 50 YR Hospitalization