STEALTHSTATION S7 SYSTEM
Report
- Report Number
- 1723170-2015-00507
- Event Type
- Malfunction
- Date Received
- April 27, 2015
- Date of Event
- March 30, 2015
- Report Date
- February 18, 2016
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- HAW
- PMA / PMN Number
- K050438
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
(B)(4).
CORRECTION: THE HARDWARE ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION HARDWARE ANOMALY TRACKING DATABASE. ADDITIONAL INFORMATION: THE SYSTEM CONTROL UNIT (SCU) WAS SENT TO THE VENDOR FOR ADDITIONAL ANALYSIS WHICH CONFIRMED THE PREVIOUSLY REPORTED FINDINGS. THE SCU WOULD NOT COMMUNICATE OR SUPPLY POWER TO THE POSITIONING SENSOR UNIT (PSU) DUE TO A FAULTY COMPONENT ON THE MAIN BOARD BEING INTERNALLY SHORTED. AS A RESULT THE EFFECTED COMPONENT WAS REPLACED FOLLOWED BY APPLICABLE TESTING TO ENSURE PROPER OPERATION.
THE SUSPECT PSU (POSITION SENSOR UNIT A.K.A. CAMERA) AND SCU (SYSTEM CONTROL UNIT) WERE RETURNED FOR ANALYSIS. RESULTS AS FOLLOWS: PSU: MANUFACTURE DATE OCT 2014. AS RECEIVED, FIRMWARE REV 14. WHEN CONNECTED TO A KNOWN GOOD SYSTEM THE PSU PERFORMED AS EXPECTED. THE PSU PASSED THE ACCURACY ASSESSMENT TESTING. NO PROBLEM FOUND. UNABLE TO DUPLICATE EVENT. SCU: MANUFACTURE DATE JUNE 2011. AS RECEIVED, FIRMWARE REV 12. WHEN CONNECTED TO A KNOWN GOOD PSU THE SCU FUNCTIONED NORMALLY AT POWER UP, BUT SOON WOULD NOT COMMUNICATE OR SUPPLY POWER TO THE PSU, ALTHOUGH THE SCU REMAINED POWERED. THE EVENT LOG HAD NOT RECORDED ANY ADVERSE EVENTS. ELECTRICAL ISSUE CAUSED EVENT.
NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. RETURN REQUESTED. REPLACEMENT SCU KIT AND PSU KIT WERE SHIPPED TO SITE. NO PARTS HAVE BEEN RECEIVED BY MANUFACTURER FOR ANALYSIS. A MEDTRONIC REPRESENTATIVE PERFORMED A NAVIGATION SYSTEM CHECK-OUT, SOFTWARE AND INSTRUMENTS AREAS PASSED. HARDWARE TESTING FOUND THE PSU WAS NOT RECEIVING POWER SUPPLY, SCU POWER CONNECTION FAILURE. POSITIONING SENSOR UNIT (PSU) WAS NOT REPLACED, IT WAS RETURNED, UNUSED TO MANUFACTURER. POLARIS SPECTRA SYSTEM CONTROL UNIT (SCU) WAS REPLACED. ISSUE RESOLVED. SYSTEM PERFORMED AS INTENDED. NO FURTHER ISSUES HAVE BEEN REPORTED.
A SITE PHYSICIAN REPORTED A NAVIGATION SYSTEM THAT WAS INCAPABLE OF TRACKING INSTRUMENTS. THE POSITIONING SENSOR UNIT (PSU) DID NOT RECEIVE ANY POWER SUPPLY, THE LEDS WERE OFF. IN TROUBLE-SHOOTING, THE POLARIS SPECTRA SYSTEM CONTROL UNIT (SCU) WAS POWERED DOWN. WHEN TOUCHING THE SCU'S POWER CABLE, IT TURNED ON, SUGGESTING A DISCONNECTION CAUSING THE ISSUE OR AN INTERMITTENT SCU POWER CONNECTOR. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 275772 | STEALTHSTATION S7 SYSTEM | NEUROLOGICAL STEREOTAXIC INSTRUMENT | HAW | MEDTRONIC NAVIGATION, INC. | S7 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |