TERMINAL CUP IMPACTOR FOR METAL BACK CC
Report
- Report Number
- 3005180920-2015-00065
- Event Type
- Malfunction
- Date Received
- April 22, 2015
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LXH
- Report Source
- Manufacturer report
Narratives
BATCH REVIEW PERFORMED ON 22 APRIL 2015: LOT 095874: (B)(4) TERMINALS PRODUCED AND RELEASED ON 30 NOVEMBER 2009. NO ANOMALIES FOUND RELATED TO THE ISSUE. LOT 132221: (B)(4) SHELLS PRODUCED AND RELEASED ON 05 AUGUST 2013. NO ANOMALIES FOUND RELATED TO THE PRODUCT. TO DATE, (B)(4) SHELLS OF THE LOT HAVE BEEN ALREADY SOLD. ON 22ND APRIL THE R AND D PROJECT MANAGER MADE THE FOLLOWING ANALYSIS UPON INSPECTION OF THE RETRIEVED ITEMS: THE DEVICES COULD NOT BE SEPARATED MANUALLY, THEREFORE A CLAMP WAS USED IN ORDER TO DISENGAGE THE TERMINAL FROM THE METAL BACK. AFTER RELEASE, IT WAS NOTICED THAT THE CONTACT SURFACE OF THE METAL BACK IS SEVERELY RUINED DUE TO THE FRICTION BETWEEN THE TERMINAL AND THE METAL BACK. THE TERMINAL GOT STUCK IN THE METAL BACK PROBABLY DUE THE HIGH FRICTION BETWEEN THE TWO DEVICES DURING IMPACTION LEADING TO TA COLD WELDING EVENT.
(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 268050 | TERMINAL CUP IMPACTOR FOR METAL BACK CC | SURGICAL INSTRUMETN FOR HIP | LXH | MEDACTA INTERNATIONAL SA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |