FDA Adverse Event Malfunction Summary report: N

IPC® HANDPIECE ATTACHMENT - INDIGO¿ DRILL

MDR report key: 4728035 · Received April 24, 2015

Report

Report Number
1625507-2015-00018
Event Type
Malfunction
Date Received
April 24, 2015
Date of Event
March 23, 2015
Report Date
March 31, 2015
Manufacturer
MDT POWERED SURGICAL SOLUTIONS
Product Code
HBE
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT ANALYSIS WAS COMPLETED ON MAY 13, 2015. THE PRODUCT ANALYSIS INDICATES THAT THE INITIAL INSPECTION OF THE DEVICE SHOWED NO SIGNIFICANT PHYSICAL DAMAGE. THE MOTOR WAS CONNECTED TO THE CONSOLE AND OPERATED AT DEFAULT SPEED WITH THE STRAIGHT AND ANGLED ATTACHMENTS. THE DRILL FUNCTIONED AND SOUNDED NORMAL WITH THE STRAIGHT ATTACHMENT (1845010). HOWEVER, THE ANGLED ATTACHMENT (1845020) SOUNDED LOUD AND THE BUR WAS DIFFICULT TO LOCK IN. THE SERVICE REPORT FOR ANGLED ATTACHMENT (1845020) INDICATES THAT THE BEARINGS AND GEARS WERE CORRODED. THE ATTACHMENT GOT TO 126.3 DEGREES F IN 5 MINUTES. THE REQUIRED COMPONENTS WERE REPLACED. RESULTS CODE: DEGRADATION PROBLEM. RESULTS: THERMAL PROBLEM. CONCLUSION: DEVICE REPAIRED AND RETURNED.

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: 1845000 - DRILL 1845000 INDIGO HIGH SPEED OTOLOGIC; MANUFACTURE DATE ¿ MAY 8, 2014; SERIAL NUMBER ¿ (B)(4); LOT ¿ 208309894; 510K - K081475 1845010 - ATTACHMENT 1845010 INDIGO OTOL STRAIGHT; MANUFACTURE DATE ¿ AUGUST 21, 2013; SERIAL NUMBER ¿ (B)(4); LOT ¿ 207155377; 510K - K081475. THE HANDPIECE, STRAIGHT ATTACHMENT, AND ANGLED ATTACHMENT WERE RETURNED. SERVICE AND REPAIR DID NOT RECEIVE AN O-RING WITH THE HANDPIECE OR THE ATTACHMENTS. 1845020: THE PRE-REPAIR TEMPERATURE TEST RESULTS FOR THE ANGLED ATTACHMENT WERE AS FOLLOWS: 198 DEGREES F AT THE NOSE, 112 DEGREES F AT THE MIDDLE, AND 99 DEGREES F AT THE REAR. SERVICE AND REPAIR WAS ABLE TO VERIFY THAT THE ANGLED ATTACHMENT DID GET HOT. THE ANGLED ATTACHMENT HAS BEEN FORWARDED TO MEDTRONIC PSS FOR ANALYSIS AND REPAIR. 1845000: THE SERVICE REPORT INDICATES THAT THE CUSTOMER'S COMPLAINT COULD NOT BE VERIFIED. THE PRE-REPAIR TEMPERATURE RESULTS WERE AS FOLLOWS: 81 DEGREES F AT THE NOSE, 79 DEGREES F AT THE MIDDLE, AND 78 DEGREES F AT THE REAR. THERE WAS NO FAULT FOUND WITH THE HANDPIECE. IT WAS CLEANED, TESTED, AND PASSED TO MANUFACTURING SPECIFICATIONS. 1845010: THE SERVICE REPORT INDICATES THAT THE CUSTOMER'S COMPLAINT COULD NOT BE VERIFIED. THE PRE-REPAIR TEMPERATURE RESULTS WERE AS FOLLOWS: 76 DEGREES F AT THE NOSE, 74 DEGREES F AT THE MIDDLE, AND 73 DEGREES F AT THE REAR. THERE WAS NO FAULT FOUND WITH THE HANDPIECE ATTACHMENT. IT WAS CLEANED, TESTED, AND PASSED TO MANUFACTURING SPECIFICATIONS. THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT INTRAOPERATIVE, THE HANDPIECE WAS GETTING HOT. THEY HAD TO USE A DAMP TOWEL TO WRAP AROUND THE DRILL WHILE BEING USED. PART WAY THROUGH THE PROCEDURE, THE DOCTOR WAS FRUSTRATED AND SWITCHED OUT THE HANDPIECE. IT WAS ALSO REPORTED THAT AN O-RING FELL OUT OF THE DRILL WHEN A DRILL BIT (BUR/ATTACHMENT) WAS BEING SWITCHED. THE HANDPIECE, STRAIGHT ATTACHMENT, AND ANGLED ATTACHMENT WERE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
274383 IPC® HANDPIECE ATTACHMENT - INDIGO¿ DRILL DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MDT POWERED SURGICAL SOLUTIONS 1845020 207305240

Patients

Seq Age Sex Outcome Treatment
1