FDA Adverse Event Death Summary report: N

CONCHATERM NEPTUNE HEATED HUMIDIFIER

MDR report key: 4726797 · Received April 24, 2015

Report

Report Number
4726797
Event Type
Death
Date Received
April 24, 2015
Date of Event
February 6, 2015
Report Date
March 13, 2015
Manufacturer
TELEFLEX MEDICAL INC
Product Code
BTT
Report Source
User Facility report
Reporter Location
NE, US
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Description of Event or Problem · 1

VENTILATOR SET UP ON (B)(6) 2015 FOR GESTATION (B)(6) INFANT WHO WAS BORN IN THE CAR WITH TWIN ENROUTE TO THE HOSPITAL. ON (B)(6) 205 BABY STARTED TO EXHIBIT SOME RESPIRATORY DISTRESS. BABY REINTUBATED MUCUS FOUND ON END OF TUBE. AFTER INTUBATION ATTEMPTED SUCTION OF BABY AND RED BLOOD NOTED. BABY DIED. ULTRASOUND OF HEAD ALSO NOTED BILATERAL GERMINAL MATRIX HEMORRHAGES. UPON CLEANING UP OF EQUIPMENT RT NOTICED WATER BOTTLE FOR ACTIVE HUMIDITY CIRCUIT WAS NOT SPIKE CAUSING THE EDEMA TO BE DRY. THEREFORE NO HUMIDITY SPIKE HAD GONE IN BUT NOT PUNCTURE ONE AND WATER BOTTLE. TEMPERATURE ALARM DID NOT SOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273111 CONCHATERM NEPTUNE HEATED HUMIDIFIER HUMIDIFIER SYSTEM BTT TELEFLEX MEDICAL INC

Patients

Seq Age Sex Outcome Treatment
1 1 DA Death