FDA Adverse Event Other Summary report: N

ECLIPSE

MDR report key: 472206 · Received July 11, 2003

Report

Report Number
2914292-2003-00004
Event Type
Other
Date Received
July 11, 2003
Report Date
July 10, 2003
Manufacturer
VARIAN MEDICAL SYSTEMS
Product Code
MUJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

A COMPUTER SYSTEM CRASH OCCURRED IN A HOSP. SUBSEQUENT TO THE CRASH, A CORRUPTED FILE WAS USED IN CALCULATING FIELDS HAVING ENHANCED DYNAMIC WEDGE RESULTING IN UNEXPECTED DOSE DISTRIBUTION. IT HAS BEEN REPORTED THAT MISTREATMENTS HAVE OCCURRED, HOWEVER THE EXTENT HAS NOT BEEN DETERMINED. THIS MDR IS FILED BECAUSE SERIOUS INJURY CANNOT BE RULED OUT AS THIS TIME. THIS EVENT WAS REPORTED IN JUNE 2003.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECLIPSE TREATMENT PLANNING SYSTEM MUJ VARIAN MEDICAL SYSTEMS VERSION 6.5 *

Patients

Seq Age Sex Outcome Treatment
1 NO INFO Other