FDA Adverse Event
Other
Summary report: N
ECLIPSE
MDR report key: 472206
·
Received July 11, 2003
Report
- Report Number
- 2914292-2003-00004
- Event Type
- Other
- Date Received
- July 11, 2003
- Report Date
- July 10, 2003
- Manufacturer
- VARIAN MEDICAL SYSTEMS
- Product Code
- MUJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- INVALID DATA
Narratives
Description of Event or Problem · 1
A COMPUTER SYSTEM CRASH OCCURRED IN A HOSP. SUBSEQUENT TO THE CRASH, A CORRUPTED FILE WAS USED IN CALCULATING FIELDS HAVING ENHANCED DYNAMIC WEDGE RESULTING IN UNEXPECTED DOSE DISTRIBUTION. IT HAS BEEN REPORTED THAT MISTREATMENTS HAVE OCCURRED, HOWEVER THE EXTENT HAS NOT BEEN DETERMINED. THIS MDR IS FILED BECAUSE SERIOUS INJURY CANNOT BE RULED OUT AS THIS TIME. THIS EVENT WAS REPORTED IN JUNE 2003.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECLIPSE | TREATMENT PLANNING SYSTEM | MUJ | VARIAN MEDICAL SYSTEMS | VERSION 6.5 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO | Other |