FDA Adverse Event Malfunction Summary report: N

SKIN STAPLE EXTRACTOR

MDR report key: 4717909 · Received April 21, 2015

Report

Report Number
1060680-2015-00014
Event Type
Malfunction
Date Received
April 21, 2015
Date of Event
March 10, 2015
Report Date
April 17, 2015
Manufacturer
DEROYAL INDUSTRIES
Product Code
KDD
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FINDINGS: THE QFI REPORT WAS REVIEWED TO OBTAIN THE SALES AND SIMILAR COMPLAINT INFORMATION. DEROYAL HAS SOLD (B)(4) CASES OF THE FINISHED GOOD FROM 2013 TO 2015. THERE HAVE BEEN NO PREVIOUS REPORTS OF THIS NATURE PRIOR TO THE CUSTOMER FILING TWO COMPLAINTS. RAW MATERIAL (B)(4) LOT NUMBER 1195857 WAS UTILIZED WITHIN THE FINISHED GOOD AND LOT NUMBER REPORTED. THE DEVICE IN USE DURING THE REPORTED INCIDENT WAS NOT RETURNED. HOWEVER, REPRESENTATIVE SAMPLES WERE RECEIVED FROM THE REPORTING CUSTOMER AND CONSISTED OF FOUR EACH OF THE DEVICE SEALED WITHIN THE PACKAGE. LOT NUMBER 36427356 WAS RETURNED AND NOT THE LOT NUMBER REPORTED. LOT MAPPING WAS PERFORMED AND IT WAS IDENTIFIED THAT ONE OF THE RAW MATERIAL LOT NUMBERS UTILIZED WAS THE SAME AS THE REPORTED LOT NUMBER: 36427356, (B)(4), 1195857; 36427356, (B)(4), 1135466. THE SAMPLES WERE FUNCTIONALLY TESTED BY REMOVING STAPLES FROM A BOX OF FOAM. ALL OF THE SKIN STAPLE REMOVERS PERFORMED AS INTENDED WITH NO IDENTIFIABLE ISSUES. A SCAR WAS ISSUED TO (B)(4)., THE RAW MATERIAL SUPPLIER, WITH A DUE DATE OF 04/20/2015. AS OF THE DATE OF THIS REPORT, THE SCAR HAS NOT BEEN RETURNED. THE INVESTIGATION IS INCOMPLETE AT THIS TIME. THIS REPORT WILL BE UPDATED WHEN NEW INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

(B)(6) MED. CTR. JUST STARTED USING THE SKIN STAPLE REMOVERS AND HAS CONCERNS ABOUT EMPLOYEES GETTING STUCK WITH THE EXTRACTED STAPLES. THE MAIN CONCERN IS THAT THE STAPLE GETS STUCK ON THE EXTRACTOR AND HAS TO BE PULLED OFF MANUALLY. IT DOES NOT FALL OFF WITHOUT ASSISTANCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262026 SKIN STAPLE EXTRACTOR KIT, SURGICAL INSTRUMENT, DISPOSABLE KDD DEROYAL INDUSTRIES 37840185

Patients

Seq Age Sex Outcome Treatment
1