FDA Adverse Event Malfunction Summary report: N

TERMINAL CUP IMPACTOR FOR METAL BACK CC

MDR report key: 4707571 · Received April 9, 2015

Report

Report Number
3005180920-2015-00052
Event Type
Malfunction
Date Received
April 9, 2015
Date of Event
March 10, 2015
Report Date
April 9, 2015
Manufacturer
MEDACTA INTERNATIONAL, SA
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 07/02/2015 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION COLLECTED DURING THE INVESTIGATION, ALREADY REPORTED IN THE INITIAL REPORT AND IN THE FIRST FOLLOW UP. ON 07/09/2015 THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: ON 04/20/2015 THE RETRIEVED ITEMS HAVE BEEN DISASSEMBLED AND THE LOT OF THE INSTRUMENT HAS BEEN DETECTED: 1213128. ON THIS DATE, THE R AND D PROJECT MANAGER ANALYZED THE ITEMS WITH THE FOLLOWING OUTCOMES: THE TERMINAL WAS BLOCKED INTO THE CUP BECAUSE THE THREAD WAS DAMAGED LIKELY DUE TO A HIT. THE DAMAGED THREAD COMPROMISES THE COUPLING WITH THE CUP.

Additional Manufacturer Narrative · 1

THE LOT OF THE INSTRUMENT IS UNK AND WE ARE WAITING FOR THE RETRIEVED ITEMS TO MAKE IT AVAILABLE. BATCH REVIEW PERFORMED ON 04/09/2015 ON THE LOT OF THE SHELL: LOT 145994: (B)(4) SHELLS PRODUCED AND RELEASED ON 11/17/2014. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) SHELLS OF THE LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR ISSUE.

Description of Event or Problem · 1

IMP REF#: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
233733 TERMINAL CUP IMPACTOR FOR METAL BACK CC SURGICAL INSTRUMENT FOR HIP LXH MEDACTA INTERNATIONAL, SA

Patients

Seq Age Sex Outcome Treatment
1 UNK Other VERSAFITCUP CC TRIO ACETABULAR SHELL:| K103352, CODE: 01.26.45.0058, LOT: 145994