TERMINAL CUP IMPACTOR FOR METAL BACK CC
Report
- Report Number
- 3005180920-2015-00052
- Event Type
- Malfunction
- Date Received
- April 9, 2015
- Date of Event
- March 10, 2015
- Report Date
- April 9, 2015
- Manufacturer
- MEDACTA INTERNATIONAL, SA
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ON 07/02/2015 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION COLLECTED DURING THE INVESTIGATION, ALREADY REPORTED IN THE INITIAL REPORT AND IN THE FIRST FOLLOW UP. ON 07/09/2015 THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.
ADDITIONAL INFORMATION: ON 04/20/2015 THE RETRIEVED ITEMS HAVE BEEN DISASSEMBLED AND THE LOT OF THE INSTRUMENT HAS BEEN DETECTED: 1213128. ON THIS DATE, THE R AND D PROJECT MANAGER ANALYZED THE ITEMS WITH THE FOLLOWING OUTCOMES: THE TERMINAL WAS BLOCKED INTO THE CUP BECAUSE THE THREAD WAS DAMAGED LIKELY DUE TO A HIT. THE DAMAGED THREAD COMPROMISES THE COUPLING WITH THE CUP.
THE LOT OF THE INSTRUMENT IS UNK AND WE ARE WAITING FOR THE RETRIEVED ITEMS TO MAKE IT AVAILABLE. BATCH REVIEW PERFORMED ON 04/09/2015 ON THE LOT OF THE SHELL: LOT 145994: (B)(4) SHELLS PRODUCED AND RELEASED ON 11/17/2014. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) SHELLS OF THE LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR ISSUE.
IMP REF#: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 233733 | TERMINAL CUP IMPACTOR FOR METAL BACK CC | SURGICAL INSTRUMENT FOR HIP | LXH | MEDACTA INTERNATIONAL, SA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other | VERSAFITCUP CC TRIO ACETABULAR SHELL:| K103352, CODE: 01.26.45.0058, LOT: 145994 |