TRUERESULT
Report
- Report Number
- 1052693-2015-00552
- Event Type
- Malfunction
- Date Received
- April 15, 2015
- Date of Event
- March 23, 2015
- Report Date
- April 15, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K051147
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). RETURNED METER AND TEST STRIPS EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION. USER HAD INACCURATE REFERENCE.
CONSUMER COMPLAINT OF HIGH BLOOD RESULTS. EXPECTED BLOOD GLUCOSE RESULTS NOT FASTING IS ABOUT 155 MG/DL. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. BACK TO BACK BLOOD TEST PERFORMED DURING CALL NOT FASTING ((B)(6) 2015; 4:21 PM) WITH RESULTS OF 192 MG/DL AND 212 MG/DL. VERIFIED STORAGE OF TEST STRIPS IS WITHIN SPECIFICATION. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 04/30/2017 AND OPEN VIAL DATE IS ONE MONTH. RECALL TEST RESULTS PERFORMED FASTING FROM METER MEMORY: MEMORY 1:289MG/DL (B)(6) 2015 04:24PM FASTING: YES. MEMORY 2:325MG/DL (B)(6) 2015 04:24PM FASTING: YES. MEMORY 3:269MG/DL (B)(6) 2015 04:24PM FASTING: YES. MEMORY 4:275MG/DL (B)(6) 2015 04:24PM FASTING: NO. MEMORY 5:280MG/DL (B)(6) 2015 04:24PM FASTING: NO. ADVERSE EVENT NOT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 252564 | TRUERESULT | BLODO GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | SIDEKICK | SR4549 |