TRUEBALANCE
Report
- Report Number
- 1052693-2015-00549
- Event Type
- Malfunction
- Date Received
- April 15, 2015
- Date of Event
- March 23, 2015
- Report Date
- April 15, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B)(4). RETURNED METER AND TEST STRIPS EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION. USER HAD INACCURATE REFERENCE.
CONSUMER COMPLAINT OF HIGH BLOOD RESULTS. CUSTOMER STATES THAT SHE FEELS WELL AND REQUIRED NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 120-130 MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 05/31/2016. CUSTOMER CONFIRMS THE STRIPS WERE STORED PROPERLY AND WERE FIRST OPENED 03/22/2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 255 MG/DL AND 327MG/DL NOT FASTING, CUSTOMER ATE 1 HOUR AGO. REVIEWED METER MEMORY: 203MG/DL, (B)(6) 2015, 03:56:00 PM, FASTING: YES. 221MG/DL, (B)(6) 2015, 03:54:00 PM, FASTING: YES. 281MG/DL, (B)(6) 2015, 03:51:00 PM, FASTING: YES. 301MG/DL, (B)(6) 2015, 03:48:00 PM, FASTING: YES. 230MG/DL, (B)(6) 2015, 03:40:00 PM, FASTING: YES. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 252548 | TRUEBALANCE | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUEBALANCE | BR4514 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |